Drain Complication - Fresenius Medical Care Liberty Select User Manual

Peritoneal dialysis cycler
Table of Contents

Advertisement

Message
Meaning
Drain
Your cycler checks
complication
halfway through a
encountered.
Drain cycle to see if
Please check
you have drained at
patient and drain
least 35% (or 42.5%
lines.
if Drain exit %
option is 85%) of
your expected drain
volume. If not, your
cycler will beep to
alert you of a low
drain rate. If you
still haven't drained
at least 70% (or
85% if Drain exit %
option is 85%) of
your previous Fill by
the end of the
Drain, your cycler
will alert you again
that you are having
drain complications.
Drain Line is
Your cycler is
blocked. Please
having trouble
check the drain
pumping used
line.
dialysate to the
drain.
EEPROM Alarm:
During startup, the
EEPROM Error!
cycler cannot
properly read the
EEPROM memory
chip.
Liberty Select Cycler User's Guide
Action Required
Check the Drain Line (Yellow
clamp) is not kinked, bent, or
clamped.
Check that your patient line
(blue clamp) is not kinked or
clamped.
Check for fibrin or clots in the
drain line. If so, contact your PD
nurse for instructions on what to
do.
Change positions to move fluid
within your abdomen.
Check that the Drain Line is
properly placed in drainage
receptacle such as a toilet, sink
or bathtub.
Check position of drain bags
and make sure they are flat and
not on top of each other.
Are you constipated? If so,
contact your PD nurse for
instructions on what to do.
Check that your catheter
extension set is adequately
secured.
Press the OK key to retry.
Make sure that the Drain Line
(Yellow clamp) is properly
connected, the clamps are open
and there is no kink in the line.
Press the OK key to retry.
Record error code.
Aseptically disconnect yourself
from cycler if connected.
Turn off cycler and call service.
P/N 480145 Rev A
231