1 Touch STAT drain
If you choose to perform a STAT drain before starting to
disconnect yourself, touch the STAT drain button. Consult
with your PD nurse if you are feeling full. See page 184 for
more information.
2 Disconnect yourself
After your last Drain or Fill of your treatment, your
treatment will be complete for the day. When your cycler
reaches this screen, you may disconnect.
To disconnect from your cycler:
Mask, then wash or sanitize hands as instructed
by your PD nurse.
Insert the stay•safe PIN connector connected to
your catheter extension set into the stay•safe
organizer clip below the control panel
Insert a new stay•safe cap on the left side of the
organizer
Turn the blue button on the PIN connector ¼ turn
clockwise and push it in to make sure that the
PIN has been inserted into the PIN connector
Close patient line (blue clamp)
Close catheter extension set clamp
Use aseptic technique to carefully disconnect
your catheter extension set from your patient line
and cap off with a new sterile stay•safe cap
Check that the new sterile stay•safe cap is
securely attached to your catheter extension set
before removing it from the stay•safe organizer
Liberty Select Cycler User's Guide
Note: An additional drain is a planned drain. A
STAT drain is an unplanned drain.
Warning: You must use aseptic technique as
directed by your PD nurse to prevent infection.
Warning: Use a new, sterile stay•safe PIN
connector and cap every time you disconnect from
your cycler. Be sure to insert the new, sterile cap
into the stay•safe New Cap Holder before you
disconnect.