Message
Supply Lines are
blocked. Please
check the supply
lines.
System Error
System Warning
Treatment
Alarm: Flowpath
error!
Treatment
Alarm: Pause
time error!
Treatment
Alarm: Pump A
vs B error!
238
Meaning
Your cycler is
having trouble
pumping out of the
solution bags.
An unrecoverable
error has occurred.
An unexpected
error has occurred.
An unrecoverable
error has occurred.
An unrecoverable
error has occurred.
An unrecoverable
error has occurred.
P/N 480145 Rev A
Action Required
Make sure that the solution lines
(White and Green clamps) are
properly connected, the clamps
are open and there is no kink in
the lines.
Make sure the cones are all
broken.
Make sure that the bags are not
empty.
Press the OK key to retry.
Record error code.
Aseptically disconnect yourself
from cycler if connected.
Turn off cycler and call service.
Press the OK key to retry.
Press STOP to mute.
Record error code.
Insert PIN.
Clamp lines.
Insert patient connector into
stay.safe Organizer.
Aseptically disconnect from
cycler.
Turn off cycler and call service.
Press STOP to mute.
Record error code.
Insert PIN.
Clamp lines.
Insert patient connector into
stay.safe Organizer.
Aseptically disconnect from
cycler.
Turn off cycler and call service.
Press STOP to mute.
Record error code.
Insert PIN.
Clamp lines.
Insert patient connector into
stay.safe Organizer.
Aseptically disconnect from
cycler.
Turn off cycler and call service.