Message
Meaning
Please wait for
The Drain phase is
solution to cool
finished but the
down
heater bag solution
temperature is
above 38°C. The
cycler will not allow
Fill to begin until the
temperature drops
below 38°C.
Please wait for
The Drain phase is
solution to warm
finished but the
up
heater bag solution
temperature is still
below 25°C. The
cycler will not allow
Fill to begin until the
temperature
reaches 33°C.
Pneumatic
The cycler has
Alarm: Balloon
failed due to a
Valve Leak!
possible balloon
valve air pressure
leak during
treatment.
Pneumatic
An air pressure leak
Alarm: Pressure
has been detected
leak!
and the cycler
cannot recover.
Pneumatic
The cycler's
Warning:
pneumatic pressure
Pressure reading
is reading too high.
high!
Pneumatic
The cycler's
Warning:
pneumatic pressure
Pressure reading
is reading too low.
low!
Liberty Select Cycler User's Guide
Action Required
Wait for the heater bag
temperature to decrease. If this
message occurs frequently,
check for external heating
sources and make sure that the
room temperature is cooler than
92°F (33°C).
Wait for the heater bag
temperature to increase. If this
message occurs frequently,
check for external cooling
sources: Be sure there are no
open windows, drafts, or fans or
air conditioners blowing on your
cycler.
Press STOP to mute.
Record error code.
Insert PIN.
Clamp lines.
Insert patient connector into
stay.safe Organizer.
Aseptically disconnect from
cycler.
Turn off cycler and call service.
Press STOP to mute.
Record error code.
Insert PIN.
Clamp lines.
Insert patient connector into
stay.safe Organizer.
Aseptically disconnect from
cycler.
Turn off cycler and call service.
Press the OK key to retry.
Press the OK key to retry.