Fresenius Medical Care Liberty Select User Manual page 253

Peritoneal dialysis cycler
Table of Contents

Advertisement

Electro-magnetic
compatibility
Electrical Safety (UL 60601-1)
Protection against
electric shock
IP Rating
Mode of
Operation
Equipment not suitable for use in the presence of a flammable anesthetic
mixture with air or with oxygen or nitrous oxide.
User Interface
Language
Touch Screen
Keys
Hardware Settings
Alarm Loudness
Key Loudness
Screen Blank-ing
Delay
Screen Brightness 1-10
Solution
Bag Volumes
Initial Fill Volume
Range
Temperature
Maximum Total
Volume
Drain Time
Fill Volume
Pause Volume
Last Fill Volume
Liberty Select Cycler User's Guide
See EMC Declaration on page 254
Type: Class I equipment
Degree: Type BF Applied Part
IPX0. Meets the requirements of IEC 60601-2-39 (Clause
44.3) regarding degree of protection against ingress of
water.
Continuous Operation
The operating screens may be set to either English (USA),
French (Canada), or Spanish (Mexico)
LCD display reacts to fingertip pressure
Up , Down , STOP, OK
1-20 (50-90 dB)
0-20
10 minutes
2000, 3000, 5000, 6000 (v2.8.4 software or later) mL
500 to 4000 mL
Fixed at 37°C 1°C (98.6°F 1.8°F)
Minimum 25°C bag temperature in order to start treatment
30,000 mL
20-59 minutes
500-4000 mL
0-4000 mL
0-4000 mL
P/N 480145 Rev A
251