Fresenius Medical Care Liberty Select User Manual page 239

Peritoneal dialysis cycler
Table of Contents

Advertisement

Message
Meaning
Self Test
The self test has
Warning:
failed due to
Pressure out of
pressures being out
range!
of range.
Self Test
The self test has
Warning:
failed due to low
Pressure
pressures.
production low!
Self Test
The self test has
Warning:
failed due to a
Thermistor
temperature sensor
reading out of
being out of range.
range!
Self Test
The self test has
Warning:
failed due to a
Vacuum leak!
possible vacuum
leak.
Self Test
The self test has
Warning:
failed due to
Vacuum out of
vacuum pressures
range!
being out of range.
Self Test
The self test has
Warning:
failed due to low
Vacuum
vacuum pressures.
production low!
Sensor Warning:
A sensor is reading
Invalid sensor
incorrectly.
readings!
Solution is
The solution
warming. Press
temperature is cool
OK key to start
(25°C/77°F). The
fill...
user must choose
whether or not to
wait for it to reach
the set temperature.
Liberty Select Cycler User's Guide
Action Required
Press the OK key to retry.
Press the OK key to retry.
Make sure nothing is on top of
heater tray.
Make sure cycler has cooled
down.
Make sure cycler is not placed
near a heat source.
Press the OK key to retry.
Press the OK key to retry.
Press the OK key to retry.
Press the OK key to retry.
Press the OK key to retry.
To continue Fill with cool
solution, press the OK key;
otherwise, wait until your
solution has reached the set
temperature to continue. When
the set temperature has been
reached, it will automatically
start the Fill.
P/N 480145 Rev A
237