Error Messages Listed Alphabetically - Fresenius Medical Care Liberty Select User Manual

Peritoneal dialysis cycler
Table of Contents

Advertisement

Error messages listed alphabetically

Message
Battery RAM
Alarm: Battery
RAM Error!
Bypass not
allowed!
CRC Alarm:
CRC Error!
230
Note: Press the STOP key to silence an audible warning or
alarm.
Meaning
The battery RAM is
corrupted.
The operator has
tried to bypass the
current phase.
The CRC is
corrupted.
P/N 480145 Rev A
Action Required
Record error code.
Aseptically disconnect yourself
from cycler if connected.
Turn off cycler and call service.
No action is required. You
cannot bypass while the heater
bag is filling, if it does not have
enough solution for the next fill.
If your cycler detects that an
overfill situation could occur
(150% = next Fill + amount not
drained from previous Fills), you
cannot bypass and the machine
will continue to attempt to drain
you until it is safe to move on to
the next Fill.
Record error code.
Aseptically disconnect yourself
from cycler if connected.
Turn off cycler and call service.
Liberty Select Cycler User's Guide