Pressures
Generated
Therapy
Total Therapy
Time
Number of Fills
Dwell Time
Pauses
Safety Systems
Air Infusion
Protection
Dialysate Solution
Overfill Protection
Over Temperature
Protection
Options
Accessories
IQdrive (P/N 180202)
252
-250 mbar to 650 mbar (pressures that can be generated to
assist the transfer of dialysate solution to and/or from the
peritoneal cavity of the patient)
10 minutes to 47 hours, 59 minutes
1-99 (Exchanges)
Variable from 0 minutes to 9 hours, 59 minutes
Variable from 0 to 1 less than or equal to # of Fills
programmed
Controlled by software, the cycler measures air volume in
three consecutive strokes coming from source. Any air
volume over the set limits (see below) is pumped to drain
and M31 (Air Detected in Cassette) warning is displayed.
Air Detector Sensitivity Limits
Controlled by software. Drain bypass is prevented when
overfill is 1.5 or 150% of prescribed fill (150% = next fill +
amount of previous fill not yet drained)
Over temperature protected by two thermostats opening at
104° ± 4°F and 120° ± 4°F
Add diurnal UF
Language
USB enabled
Flow alert
P/N 480145 Rev A
Fill: air volume coming from heater bag ≥ 3 mL
Drain: air volume coming from patient's abdominal
cavity ≥ 3 mL
Dwell: If air volume coming from supply bag ≥ 6 mL
Last bag option
SI Units
Add. drain
Flow alert sound
Liberty Select Cycler User's Guide
Drain exit %
Personal data
Add. drain alert