Fresenius Medical Care Liberty Select User Manual page 231

Peritoneal dialysis cycler
Table of Contents

Advertisement

Code
Message
S_002
CRC Alarm: CRC
Error!
S_003
Battery RAM Alarm:
Battery RAM Error!
S_004
EEPROM Alarm:
EEPROM Error!
S_005
System Error
Liberty Select Cycler User's Guide
Meaning
The CRC is
corrupted.
The battery RAM
is corrupted.
During startup, the
cycler cannot
properly read the
EEPROM memory
chip.
An unrecoverable
error has
occurred.
P/N 480145 Rev A
Action Required
Record error code.
Aseptically
disconnect
yourself from
cycler if
connected.
Turn off cycler and
call service.
Record error code.
Aseptically
disconnect
yourself from
cycler if
connected.
Turn off cycler and
call service.
Record error code.
Aseptically
disconnect
yourself from
cycler if
connected.
Turn off cycler and
call service.
Record error code.
Aseptically
disconnect
yourself from
cycler if
connected.
Turn off cycler and
call service.