Fresenius Medical Care Liberty Select User Manual

Fresenius Medical Care Liberty Select User Manual

Peritoneal dialysis cycler
Table of Contents

Advertisement

Quick Links

User's Guide
Select

Advertisement

Table of Contents
loading

Summary of Contents for Fresenius Medical Care Liberty Select

  • Page 1 User’s Guide Select...
  • Page 3 Caution: Frequency, duration, and treatment settings are to be determined by the prescribing physician. Indications for Use: The Fresenius Liberty Select Cycler is indicated for acute and chronic peritoneal dialysis. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 4: How To Use This User's Guide

    Guide This User’s Guide is a reference for you to look up information about the Liberty Select Cycler. Start with the topic you are interested in, and look for it in the Table of Contents (see page 4) or in the Index (see page 266). The Table of Contents is detailed enough that you should be able to find most topics.
  • Page 5: Table Of Contents

    How to insert stay•safe PIN connector into stay•safe organizer: 106 Exchanges How to bypass: 186 How to do a STAT drain: 184 Drain times based on Fill volume: 240 Technical Support Number: 1-800-227-2572 Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 6: Table Of Contents

    My Treatment Settings ...............6   Introduction ..................12   Requirements ................12   Purpose of the Liberty Select Cycler ..........12   Contraindications (When this machine should not be used) ..13   Peritoneal Dialysis Precautions ........... 13   About This User’s Guide .............. 14  ...
  • Page 7: Table Of Contents

    Bypassing a Phase ..............186   Using the IQdrive (Optional) ............194   Using the Wireless Modem (Optional) ........200   Moving Your Liberty Select Cycler ..........202   Warnings and Alarms ..............206   Warnings and Alarms (continued) ..........208  ...
  • Page 8: My Treatment Settings

    My Treatment Settings Use this page for your treatment values as prescribed by your doctor or practitioner. Use only one of these prescriptions for your treatment. P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 9 Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 10 P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 11 Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 12 Notes: ________________________________________________ ______________________________________________________ ______________________________________________________ ______________________________________________________ ______________________________________________________ ______________________________________________________ ______________________________________________________ ______________________________________________________ P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 13 Introduction & Overview Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 14: Introduction

    Be sure to read and follow all instructions and warnings listed in the User’s Guide. Requirements Operators like you must be trained on the Liberty Select Cycler to perform peritoneal dialysis under the direction of a physician. In addition, each operator should be: ...
  • Page 15: Contraindications (When This Machine Should Not Be Used)

     Multiple abdominal surgeries This is a guideline only. Your doctor or practitioner must make the final decision whether the Liberty Select Cycler is appropriate for you. MR (Magnetic Resonance) Unsafe: Liberty Select Cycler must not to be used in an MR environment.
  • Page 16: About This User's Guide

    About This User’s Guide This User’s Guide is a reference for properly trained home users of the Liberty Select Cycler—it does not replace formal training by your PD nurse. It cannot provide or replace clinical advice.
  • Page 17 MR environment. This symbol on the back of your cycler indicates that Interference may occur in the vicinity of the machine. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 18: Peritoneal Cycler Preventive Maintenance

    The only required maintenance is surface cleaning as stated in the User’s Guide. If problems do occur, in most cases Fresenius Medical Care will replace the cycler within 1 working day and, if applicable, arrange to pick up the used cycler. The replacement policy applies only to those cyclers that remain the property of Fresenius Medical Care North America.
  • Page 19: Incomplete Treatment

    Warning: Keep fingers clear of moving parts. Warning: The Liberty Select Cycler is not sterile equipment. Only the interior of the tubing sets are sterilized. You must use aseptic technique (masking and hand washing) as directed by your PD nurse to prevent infection.
  • Page 20 Warning: Do not allow any liquid to drip in or on your machine. Spills from your treatment can lead to serious injury or death. If spills occur, immediately contact a Fresenius Medical Care Technical Support Representative at 1-800-227-2572. Do not attempt to clean or disinfect leaks inside your cycler.
  • Page 21 Drain 0 when your cycler starts Fill 1. Drain 0 may be bypassed under certain circumstances. Please consult your PD nurse before bypassing Drain 0. See page 186 for more information. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 22 Caution: Do not place your cycler near external heating or cooling sources like heaters or air conditioners. Caution: Your Liberty Select Cycler contains an internal battery. This battery requires no periodic maintenance and you should not attempt to replace it.
  • Page 23 Note: Under normal operating conditions, the expected useful life of the Liberty Select Cycler is five years. Note: If it becomes necessary to disconnect, the set in use must have a second unused sterile PIN connector available to resume treatment.
  • Page 24: Meet Your Liberty Select Cycler

    Meet Your Liberty Select Cycler P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 25 Liberty Select Cycler Purpose Your Liberty Select Cycler helps your body to do the work of your kidneys. Your cycler pumps dialysate (a cleansing solution used to draw waste products and extra fluids out of your blood) into your empty peritoneal cavity (Fill phase).
  • Page 26 Heater Touchscreen Tray Sensor Control Panel Pump Cabinet Door Module Latch Your Liberty Select Cycler - Front View P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 27 IQdrive Slot / Power USB Port Vent Switch Power Power Cord Cord Socket Your Liberty Select Cycler - Rear View Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 28: The Control Panel

    The Control Panel Touch- screen Status Main Button Up/Down STOP Area Area Arrow Keys P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 29 (see page 190 for instructions on how to perform bypass). STOP key The STOP key stops functions on your machine. Pressing the STOP key during an audible alarm will silence the alarm for ten minutes. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 30: The Status Bar

    The Status Bar P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 31 If the Status Bar changes to red, you must stop your treatment and call a Fresenius Medical Care Technical Support Representative at 1-800-227-2572. For more information on alarms, see “Warnings and Alarms,” on page 206.
  • Page 32: Touch Screen Buttons

    Gray or light background color blue Screen -Buttons Movement -Buttons Parameter -Buttons Note: The values shown here are for example only. You must use the values prescribed by your doctor or practitioner. P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 33 Depending on the button, you may edit that parameter either directly or through another screen using other keys and buttons. Grayed-out parameter-buttons are unavailable options. Blue parameter-buttons are not editable; they display treatment settings affected by other active parameter-buttons. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 34: Navigating With Screen-Buttons

    Navigating With Screen- Buttons Screen-buttons Note: The values shown here are for example only. You must use the values prescribed by your doctor or practitioner. P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 35 My Records The My Records screen-button allows you to view your current treatment data and treatment data from past treatments. For more information, see “The My Records Screen-Button,” on page 176. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 36: The Pump Module

    The Pump Module Door Cassette Cassette Door Sensor Guide Pins Pump Plate Latch Safety Pressure Alignment Cassette Pump Cassette Valves Clamp Sensor Catch Heads Door P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 37 Caution: When the cassette is properly inserted in your machine, the domes will face outward. If you insert the cassette in your machine the wrong way, it may damage your machine. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 38 Door Latch The Door Latch locks the Pump Door. During treatment, do not attempt to unlatch the Pump Door unless instructed to do so. Warning: Keep fingers clear of moving parts. P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 39 This Page is Intentionally Blank Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 40: The Cassette And Tubing Set

    The Cassette and Tubing stay•safe PIN connector 1 stay•safe PIN connector 2 stay•safe Patient Last Solution Heater Drain Line Bags (Yellow Connector (Blue (Green (White (Red Clamp) Clamp) Clamp) Clamp) Clamp) P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 41 The Bag 4 Line has a green clamp. It connects to your last bag if needed. The last bag may be used for more solution or a different solution (with medications added or with a different glucose concentration) for your last fill. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 42 Note: Do not place items other than solution bags on the Heater Tray. Drain line (yellow clamp) The Drain Line has a yellow clamp. It connects to your drain line extension or the drain bag set. P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 43 This Page is Intentionally Blank Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 44: The Stay•Safe Organizer

    The stay•safe Organizer Insert below Control Panel New Cap stay•safe Tubing Extension Holder Organizer Clip Channel Cap Holder P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 45 Extension Cap Holder The Extension Cap Holder is the right notch of the organizer. Place the end of your catheter extension set into the right notch. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 46: The Heater Tray

    The Heater Tray Heater Tray Sensor P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 47 Caution: Do not lift your machine by the heater tray. Caution: Do not place items other than solution bags on the Heater Tray. Bag Sensor The Bag Sensor measures the temperature of the dialysate solution in the heater bag. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 48: Setting Up Your Liberty Select Cycler

    Setting up Your Liberty Select Cycler Lift by handles P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 49 Setting up your Liberty Select Cycler Your Liberty Select Cycler is easy to set up. Follow the instructions below: 1 Remove your Liberty Select Cycler from box  Pull out the blue foam.  Pull dust cover aside.  Pick up your cycler by both handles on either side.
  • Page 50 1-800-227-2572. Do not alter the power cord. Do not use the power cord if broken or damaged. Your machine must be plugged directly into the electrical outlet; do not use extension cords or power strips. P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 51 My Settings Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 52: Entering Your Prescription

    Entering Your Prescription P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 53 1 Turn power switch to ON Locate the white power switch on the back of your machine and set it to the ON position. The “Ready” screen will appear on the Touch Screen shortly. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 54: Entering Your Treatment Settings

    Entering Your Treatment Settings Note: The values shown here are for example only. You must use the values prescribed by your doctor or practitioner. P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 55 Note: If you are using the IQdrive to program your treatment, you do not need to enter your therapy information. Turn to page 76 to enter your patient, options, and hardware settings. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 56: Select Therapy Tab

    Select Therapy Tab Note: The values shown here are for example only. You must use the values prescribed by your doctor or practitioner. P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 57 See page 136 for information on the different therapy types. Warning: The values and selections shown in pictures here are for example only. You must enter the values prescribed by your doctor or practitioner; see “My Treatment Settings,” page 6. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 58: Ccpd: Treatment-Based Settings

    CCPD: Treatment-Based Settings Note: The values shown here are for example only. You must use the values prescribed by your doctor or practitioner. P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 59 Press the OK key to confirm and save your settings. 8 Touch Back Touch the Back button to return to the “My Settings” main screen. Skip to page 76 to enter your personal Patient Settings. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 60: Ccpd: Cycle-Based Settings

    CCPD: Cycle-Based Settings Note: The values shown here are for example only. You must use the values prescribed by your doctor or practitioner. P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 61 Press the OK key to confirm and save your settings. 8 Touch Back Touch the Back button to return to the “My Settings” main screen. Skip to page 76 to enter your personal Patient Settings. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 62: Ipd: Treatment-Based Settings

    IPD: Treatment-Based Settings Note: The values shown here are for example only. You must use the values prescribed by your doctor or practitioner. P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 63 Press the OK key to confirm and save your settings. 8 Touch Back Touch the Back button to return to the “My Settings” main screen. Skip to page 76 to enter your personal Patient Settings. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 64: Ipd: Cycle-Based Settings

    IPD: Cycle-Based Settings Note: The values shown here are for example only. You must use the values prescribed by your doctor or practitioner. P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 65 Press the OK key to confirm and save your settings. 8 Touch Back Touch the Back button to return to the “My Settings” main screen. Skip to page 76 to enter your personal Patient Settings. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 66: Pd+: Treatment-Based Settings

    PD+: Treatment-Based Settings Note: The values shown here are for example only. You must use the values prescribed by your doctor or practitioner. P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 67 Press the OK key to confirm and save your settings. 8 Touch Back Touch the Back button to return to the “My Settings” main screen. Skip to page 76 to enter your personal Patient Settings. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 68: Pd+: Cycle-Based Settings

    PD+: Cycle-Based Settings Note: The values shown here are for example only. You must use the values prescribed by your doctor or practitioner. P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 69 Press the OK key to confirm and save your settings. 8 Touch Back Touch the Back button to return to the “My Settings” main screen. Skip to page 76 to enter your personal Patient Settings. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 70: Tidal: Treatment-Based Settings

    Tidal: Treatment-Based Settings Note: The values shown here are for example only. You must use the values prescribed by your doctor or practitioner. P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 71 7 Press OK Press the OK key to confirm and save your settings. 8 Touch Next Touch the Next button to continue to the next screen to finish entering prescribed values. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 72 Tidal: Treatment-Based Settings (continued) Note: The values shown here are for example only. You must use the values prescribed by your doctor or practitioner. P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 73 Touch the Back button to return to the previous screen, then touch the Back button again to reach the “My Settings” main screen. Skip to page 76 to enter your personal Patient Settings. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 74: Tidal: Cycle-Based Settings

    Tidal: Cycle-Based Settings Note: The values shown here are for example only. You must use the values prescribed by your doctor or practitioner. P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 75 7 Press OK Press the OK key to confirm and save your settings. 8 Touch Next Touch the Next button to continue to the next screen to finish entering prescribed values. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 76 Tidal: Cycle-Based Settings (continued) Note: The values shown here are for example only. You must use the values prescribed by your doctor or practitioner. P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 77 Touch the Back button to return to the previous screen, then touch the Back button again to reach the “My Settings” main screen. Skip to page 76 to enter your personal Patient Settings. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 78: Patient Settings Tab

    Patient Settings Tab Note: The values shown here are for example only. You must use the values prescribed by your doctor or practitioner. P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 79 You will automatically move to the keyboard screen. Note: You only need to enter your personal patient information the first time you start up your machine. Your information will be saved for later treatments. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 80 Entering Your First Name P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 81 ID buttons on the “My Settings: Patient Settings” screen. Your ID number is assigned by your PD nurse. See the “My Treatment Settings” sheet on page 2 for your ID number. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 82 Entering Your Date of Birth Note: The values shown here are for example only. You must use the values prescribed by your doctor or practitioner. P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 83 9 Press OK When you are finished entering your birth year, press the OK key to confirm and save your date of birth. Continue on to the next page to enter your weight. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 84 Entering Your Weight Note: The values shown here are for example only. You must use the values prescribed by your doctor or practitioner. P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 85 12 Press OK When you are finished entering your weight, press the OK key to confirm and save your selection. Continue on to the next page to enter your gender. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 86 Entering Your Gender Note: The values shown here are for example only. You must use the values prescribed by your doctor or practitioner. P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 87 When you are finished entering your gender, press the OK key to confirm and save your selection. 16 Touch options Touch the options button to go to the “Options Settings” screen. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 88: Options Settings Tab

    Options Settings Tab Note: The values shown here are for example only. You must use the values prescribed by your doctor or practitioner. P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 89 When you touch a yellow button, it will turn orange. Use the Up/Down Arrow keys to set it to your (/) prescribed value. Repeat these steps to enter all of your prescribed values. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 90 3 Press OK Press the OK key to confirm and save your settings. 4 Touch hardware Touch the hardware button to go to the “Hardware Settings” screen. P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 91 This Page is Intentionally Blank Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 92 Options Settings Tab (continued) Note: The values shown here are for example only. You must use the values prescribed by your doctor or practitioner. P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 93 50%, and the last is 75%. Each series of beeps occur about one minute apart as long as the cycler detects slow or no flow. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 94 Repeat these steps to enter all of your prescribed values. 3 Press OK Press the OK key to confirm and save your settings. 4 Touch hardware Touch the hardware button to go to the “Hardware Settings” screen. P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 95 This Page is Intentionally Blank Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 96: Hardware Settings Tab

    Hardware Settings Tab P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 97 The screen will blank- out after ten minutes if screen blanking is set to YES. Touch anywhere on the screen to restore the display, when desired. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 98 This Page is Intentionally Blank P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 99 My Treatment Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 100: Pre-Treatment Checklist

    Pre-Treatment Checklist P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 101 IQdrive is inserted properly into the IQdrive slot on the back of your machine (see page 194). Note: Your Liberty Select Cycler automatically runs a self-test during startup and alerts you if a problem exists. No manual testing of your cycler is required after installation.
  • Page 102: Starting Your Treatment

    Starting Your Treatment P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 103: Treatment Setup

    Warning: Check for proper functioning of your machine before beginning your treatment. Note: This User’s Guide is only a guide for the Liberty Select Cycler screens. Follow your exact screen order on the machine and use these pictures only as a reference.
  • Page 104 Starting Treatment (continued) Note: The values shown here are for example only. You must use the values prescribed by your doctor or practitioner. P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 105 Warning: Your doctor or practitioner will prescribe your entire treatment. Using incorrect values could result in complications that may be severe. Do not use example values shown in machine pictures. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 106: Preparing

    Preparing 4 Gather Supplies P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 107 Caution: The cassette door is pressurized. Do not attempt to reopen the cassette door until your machine instructs you to. Do not touch the door sensor when loading the cassette, see page 34. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 108 PIN connector 2 stay•safe PIN stay•safe Organizer connector 1 P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 109 Notify your PD nurse if you find a fluid leak. To insert the cassette:  Make sure that the Liberty Select cycler is positioned near the front of the table or cart so that all lines can hang freely from the cassette.
  • Page 110 If using a cycler cart, make sure the tubing is hanging down freely behind the handle. o If using a nightstand, make sure the tubing is hanging down freely over the front of the nightstand. P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 111 Caution: Do NOT force the door closed. If you are having trouble closing the door, open the cassette door and make sure the cassette has been inserted correctly. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 112 Select Heater Bag P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 113 9 Touch Next Touch the Next button to continue. Your cycler will now prompt you to connect your solution and drain lines, and place your bag on the heater tray. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 114 Connect Solution Lines Heater Bag Placement Line up solution bag at corners. Do not extend bag over edge P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 115 Warning: You must properly connect your solution bags as prescribed. Warning: Do not use external heating sources not authorized by Fresenius Medical Care to warm your solution bags. Contact your PD nurse for more information. Liberty Select Cycler User’s Guide...
  • Page 116 All the prescribed lines should be connected at this point. Touch the Next button to continue. 14 Touch Back Press the Back button to go back to the previous screen to change the heater bag volume. P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 117 This Page is Intentionally Blank Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 118 Break Cones and Clamp Unused Lines 15 Initiate Solution Flow P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 119 (yellow clamp), and patient line (blue clamp) open throughout setup and treatment. 17 Touch Next Touch the Next button to continue. Your cycler will now test to see which lines are connected. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 120 Check Bags Connected Unconnected (with unclamped (or clamped line) line) Note: This picture is an example only. You must follow your prescription and the instructions of your PD nurse. P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 121 19 Touch Next When each of the connected bags is correctly displayed, touch the Next button. Your machine will begin flushing the lines. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 122 Performing Flush P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 123 Note: If you need to stop the flushing process, press the STOP key. The pump will stop and the “Flushing Stopped” screen will be displayed. Flushing will resume when you press the OK key. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 124: Priming

    Priming P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 125 21 Touch Next Touch the Next button to begin the priming process. Note: During the prime phase, the patient line (blue clamp) must be lower than the heater bag. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 126 Performing Prime P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 127 Performing Prime Your machine will now prime the lines. The screen displays a timer to count down the Priming time. The next screen will appear when the priming is complete. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 128 Priming Complete P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 129 A manual exchange is when you connect your catheter extension set to a bag of solution and perform an exchange without using your Liberty Select Cycler. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 130 Entering a Daytime Exchange Note: The values shown here are for example only. You must use the values prescribed by your doctor or practitioner. P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 131 If you do not program this information into your Liberty Select Cycler at this time, your cycler will not know that you have additional fluid in your abdomen. Do NOT bypass Drain 0. Your Liberty Select Cycler will switch to Fill 1 when you are finished draining.
  • Page 132 Confirming Your Daytime Exchange P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 133 If your daytime exchange values are listed correctly, touch the Next button to continue to the next screen. The next screen will provide instructions on how to connect to your machine. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 134: Ready For Treatment - Connecting Yourself

    Ready for Treatment - Connecting Yourself stay•safe Organizer P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 135 Press the Next button to begin your prescribed exchanges. Warning: If you turn off your cycler after you have filled the abdomen, when you restart your cycler you MUST NOT bypass Drain 0. It could result in an overfill. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 136 Warning: If you do one or more manual daytime exchanges that are not programmed into the Liberty Select Cycler, your cycler will not know that you have additional fluid in your abdomen. Do NOT bypass Drain 0. Your Liberty Select Cycler will switch to Fill 1 when you are finished draining.
  • Page 137 Exchanges Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 138: Exchanges

    Exchanges Drain Fill Dwell/Pause P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 139 The following screens are shown only to explain the different features. Follow your own machine’s setup. Your cycler will automatically take you through your prescription. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 140: Exchange: Drain 0

    Exchange: Drain 0 Ultrafiltration (UF) Expected drain volume Total volume drained Note: The values shown here are for example only. You must use the values prescribed by your doctor or practitioner. P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 141 Drain and this is the last Drain of your treatment, then turn to “Ending Treatment” on page 154. For Fills, see the Fill explanation on the next page. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 142: Exchange: Fill1

    Exchange: Fill1 Volume set Volume filled Note: The values shown here are for example only. You must use the values prescribed by your doctor or practitioner. P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 143 (page 148) or Pause (page 142) depending on your therapy. If you programmed a Last Fill, and this is the last Fill of your treatment, go to “Ending Treatment” on page 154. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 144: Exchange: Pause

    Exchange: Pause Turn ¼ turn clockwise and push in Blue Button Close Clamp P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 145 Check that the new sterile stay•safe cap is securely attached to your catheter extension set before removing it from the stay•safe organizer Your cycler will display how long you pause under ‘Pause time’. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 146 You may now leave your machine. Touch the Next button when you are ready to reconnect to your cycler. Note: The Next button is not available until your cycler has finished filling up the heater bag. P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 147 This Page is Intentionally Blank Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 148 Exchange: Pause Re-Start stay•safe PIN connector 2 stay•safe Organizer stay•safe PIN connector 1 Turn ¼ turn clockwise and push in Blue Button P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 149  Open catheter extension clamp 4 Touch Next When you touch the Next button, your machine will begin the Drain process. Turn to page 138 for more information on the Drain process. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 150: Exchange: Dwell

    Exchange: Dwell Note: The values shown here are for example only. You must use the values prescribed by your doctor or practitioner. P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 151 When your Dwell time is complete, your cycler will drain you again. See page 138 for more information on the Drain process. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 152: Additional Drain Option

    Additional Drain Option P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 153 NO option: If you do not want to perform an additional drain, touch the NO button. Your cycler will begin your Last Fill, if programmed, or it will end the treatment. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 154 This Page is Intentionally Blank P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 155 Ending Treatment Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 156: Ending Your Treatment

    Ending Your Treatment 2 (Disconnect) Turn ¼ turn clockwise and push in Blue Button Close Clamp P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 157  Check that the new sterile stay•safe cap is securely attached to your catheter extension set before removing it from the stay•safe organizer Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 158 YES (see “Options Settings” on page 86), your cycler will next go to the “Personal Data” screen. If the ‘personal data’ option is not set, turn to page 166. P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 159 This Page is Intentionally Blank Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 160 Entering Your Personal Health Data Note: The values shown here are for example only. You must use the values prescribed by your doctor or practitioner. P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 161 5 Enter values When you touch a yellow button, it will turn orange. Use the Up/Down Arrow (/) keys to enter your measured values. Repeat these steps to enter all of your information. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 162 6 Touch Next After pressing the OK key, touch the Next button to continue. If you press Next without entering your Personal Data, no information will be recorded for this treatment date. P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 163 This Page is Intentionally Blank Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 164: Your Treatment Summary

    Your Treatment Summary Note: The values shown here are for example only. You must use the values prescribed by your doctor or practitioner. P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 165 (due to a drain or fill bypass), the Dwell Time button will display the word ‘lost’ before the time. 8 Touch Next Touch the Next button when you are finished reviewing your treatment. Turn to page 166 to continue. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 166 Treatment Summary (continued) Note: The values shown here are for example only. You must use the values prescribed by your doctor or practitioner. P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 167 Touch the Back button to return to the “Treatment Summary” screen. Then, to continue to the next step on the next page, touch the Next button when on the “Treatment Summary” screen. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 168 Remove Set Screen 1 P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 169 To contain such spills, your machine should be on a spill-tolerant surface. Clean up spills immediately. 12 Touch Next Touch the Next button to continue to the next step. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 170 Remove Set Screen 2 P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 171 Caution: If your pump module has moisture in it, your cassette may have a leak. Save your tubing set and call a Fresenius Medical Care Technical Support Representative immediately at 1-800-227- 2572. You will be contacted about having the used tubing set sample picked up.
  • Page 172: Power Down Cycler

    Power Down Cycler view your health and treatment data P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 173 My Records If you would like to further review your treatment information, restart your cycler and touch the My Records screen-button. See page 176 for an explanation of the “My Records” screen. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 174 This Page is Intentionally Blank P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 175 Appendix Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 176: Appendix

    Appendix P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 177  Using the IQdrive—Turn to page 194  Moving your Liberty Select Cycler—Turn to page 202 If you get ‘lost’ in any of these screens, simply touch the Back button, if available, to go to a previous screen. You may also touch any screen-button in the Status Bar to view other screens.
  • Page 178: The My Records Screen-Button

    The My Records Screen- Button Note: The values shown here are for example only. You must use the values prescribed by your doctor or practitioner. P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 179 Your cycler saves your last 62 days of treatment data. 3 Touch Details Touch the Details button to view the treatment details for the selected date. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 180 Treatment Details Note: The values shown here are for example only. You must use the values prescribed by your doctor or practitioner. P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 181 Note: To go back and select a new treatment date, touch the Back button. 5 Touch Next Touch the Next button to view your personal health data for the selected treatment. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 182 Your Personal Health Data Note: The values shown here are for example only. You must use the values prescribed by your doctor or practitioner. P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 183 Touch the Back button to return to the “Your Personal Health Data” screen. Note: You may touch any of the screen-buttons in the Status Bar to exit the “Personal Health Data” screens. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 184: Using The Stop Key

    Using the STOP key P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 185 The word ‘stopped’ appears at the top of the screen  The words ‘press OK to continue’ are displayed at the bottom of the screen. When you are ready to continue, press the OK key. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 186: Using The Stat Drain

    Using the STAT Drain P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 187 If you choose not to perform a STAT drain, either touch the Cancel button or wait 10 seconds. Your cycler will go back to the “Stopped” screen (see page 182). Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 188: Bypassing A Phase

    Drain 0 Bypassed With Undrained Peritoneal Cavity (Scenario 2) Drain 0 Begins Fill 1 Begins with Fill 1 Fills Dwell 1 with But Phase is Undrained Prescribed Excess Volume Bypassed Peritoneal Cavity Volume P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 189 IIPV/overfill may result in serious injury or death from conditions such as: abdominal wall and/or diaphragmatic hernias, hydrothorax, heart problems including chest pain and heart failure, high blood pressure, pulmonary edema, decreased pulmonary function and Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 190 STAT Drain as instructed by your PD nurse. See page 184 for more information on how to use the STAT Drain feature. Once the fluid is completely drained from your abdomen, call your doctor, practitioner, or PD nurse. P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 191 This Page is Intentionally Blank Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 192 Note: If you accidentally reach this screen and do not want to bypass the current phase, wait a few seconds and the cycler will return to the previous screen. 1 Press and hold OK for four seconds P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 193 Overfilling can cause serious injury or death. Warning: If you do not program a manual daytime exchange into your Liberty Select Cycler, your cycler will not know that you have additional fluid in your abdomen. Do NOT bypass Drain 0. Your Liberty Select Cycler will switch to Fill 1 when you are finished draining.
  • Page 194 Your treatment may not be as effective. The letter ‘B’ will be displayed next to your treatment details in the “My Records” screen to show a bypass occurred. For more information, see “Treatment Details” on page 176. P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 195 This Page is Intentionally Blank Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 196: Using The Iqdrive (Optional)

    Using the IQdrive (Optional) IQdrive slot / USB Port IQdrive slot / USB Port View From Back of Liberty Select Cycler IQdrive P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 197 Caution: Your Liberty Select Cycler IQdrive is only for use with your Liberty Select Cycler. Do not use your IQdrive to store your own data from other sources (for example, digital photos from your computer).
  • Page 198 USB Upload Select option Note: The values shown here are for example only. You must use the values prescribed by your doctor or practitioner. P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 199 IQdrive with an updated prescription. Contact your PD nurse if you need to update your prescription on your IQdrive. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 200 The data files transferred via the IQdrive contain only patient data. The data is encrypted with a key based encryption. The encrypted data cannot be read without reapplying the encryption algorithm thereby ensuring the security of the data. P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 201 This Page is Intentionally Blank Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 202: Using The Wireless Modem (Optional)

    Using the Wireless Modem (Optional) RS-232 Port (Modem Connection) View From Back of Liberty Select Cycler P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 203 The data files transferred via the external cellular modem contain only the treatment and patient information collected and stored within the Liberty Select Cycler. The data is encrypted with a key-based encryption. Encrypted data cannot be read without reapplying the encryption algorithm, or ‘key’, thereby ensuring the security of the...
  • Page 204: Moving Your Liberty Select Cycler

    Moving Your Liberty Select Cycler P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 205 Moving your Liberty Select Cycler Your Liberty Select Cycler is designed to be portable so you may easily continue your dialysis therapy when you travel. Unplug your cycler’s power cord from the unit before moving it. Hold your cycler by its handles on either side when lifting it.
  • Page 206 This Page is Intentionally Blank P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 207 Alarms & Troubleshooting Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 208: Warnings And Alarms

    Warnings and Alarms Press STOP key to silence alarm P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 209 Warnings and Alarms Overview Your Liberty Select Cycler has an advanced system of features that work for your safety. Electronic sensors and diagnostic software constantly monitor your machine’s operation. When problems or possible problems are detected, your machine warns you with Status Bar messages and audible alarms.
  • Page 210: Warnings And Alarms (Continued)

    Warnings and Alarms (continued) Error Code P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 211  Check that the new sterile stay•safe cap is securely attached to your catheter extension set before removing it from the stay•safe organizer  Call Technical Support at 1-800-227-2572. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 212 A self test has failed after repeated retests. System Error A critical error has occurred. Treatment Alarm A Treatment Warning was not able to be corrected. Watchdog Alarm A critical software error has occurred. P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 213 This Page is Intentionally Blank Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 214: Warning: The Help Button

    Warning: The Help Button P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 215 Note: During an alarm or warning, all the screen- buttons in the Status Bar are unavailable. Note: For more information on correcting alarms, see “Troubleshooting,” page 218. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 216 Warning: Help Screens P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 217 Follow the steps on the screen to correct a warning situation. If you are unable to clear a warning by following the instructions on the screen or in this User’s Guide, call a Fresenius Medical Care Technical Support Representative at 1-800-227-2572. Warning: Do not continue your treatment until the warning situation has been corrected.
  • Page 218: Alarm: Red Status Bar

    Alarm: Red Status Bar Record Error Press STOP to mute Turn power OFF Code P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 219 Warning: If your machine alarms, do not attempt to use it for treatment. You will also be asked for your Liberty Select Cycler serial number and software version, which is found on the top of your cycler above the cassette door.
  • Page 220: Troubleshooting

    (3) times before calling your PD nurse or a Fresenius Medical Care Technical Support Representative at 1-800-227-2572. Note: Machines taken out of service should be tested and repaired by a qualified FMCNA technician. P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 221 The first table (page 221) lists every error code and the corresponding alarm or warning message. The second table (page 230) lists all messages in alphabetical order. (continued) Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 222 Be sure to record the error code so you may look up the alarm message. If you call a Fresenius Medical Care Technical Support Representative at 1-800-227-2572, they will also need this code. The error code is located below the My Treatment button: P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 223: Error Messages Listed By Code

    Record error code. If connected, insert PIN. Clamp lines. Insert patient connector into stay.safe Organizer. Aseptically disconnect from cycler. Turn off cycler and call service. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 224 An unrecoverable mute. Treatment data error has error! occurred. Record error code. Insert PIN. Clamp lines. Insert patient connector into stay.safe Organizer. Aseptically disconnect from cycler. Turn off cycler and call service. P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 225 An unrecoverable mute. Flowpath error! error has occurred. Record error code. Insert PIN. Clamp lines. Insert patient connector into stay.safe Organizer. Aseptically disconnect from cycler. Turn off cycler and call service. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 226 Self Test Warning: The self test has Press the OK key Vacuum leak! failed due to a to retry. possible vacuum leak. P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 227 Make sure cycler is not placed near a heat source. Press the OK key to retry. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 228 Record error code. pressure leak. Insert PIN. Clamp lines. Insert patient connector into stay.safe Organizer. Aseptically disconnect from cycler. Turn off cycler and call service. P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 229 An unrecoverable mute. Treatment start error has error! occurred. Record error code. Insert PIN. Clamp lines. Insert patient connector into stay.safe Organizer. Aseptically disconnect from cycler. Turn off cycler and call service. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 230 Turn off cycler and call service. Watchdog Alarm: An unrecoverable Record error code. Watchdog timer error has Aseptically Error! occurred. disconnect yourself from cycler if connected. Turn off cycler and call service. P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 231 Turn off cycler and call service. Record error code. S_005 System Error An unrecoverable error has Aseptically occurred. disconnect yourself from cycler if connected. Turn off cycler and call service. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 232: Error Messages Listed Alphabetically

    Fill. Record error code. CRC Alarm: The CRC is CRC Error! corrupted. Aseptically disconnect yourself from cycler if connected. Turn off cycler and call service. P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 233: Drain Complication

    EEPROM Alarm: During startup, the Record error code. EEPROM Error! cycler cannot Aseptically disconnect yourself properly read the from cycler if connected. EEPROM memory Turn off cycler and call service. chip. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 234 Make sure cycler is not placed near a heat source. Press the OK key to retry. P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 235 “flow alert”. You do Check that your catheter not need to press extension set is adequately STOP key at this secured. time. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 236: Patient Line Is Blocked

    Cassette Door. treatment has Make sure the door is closed started. Or your securely. cassette door is Press the OK key to retry. open when it should be closed. P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 237 The cycler’s Warning: pneumatic pressure Pressure reading is reading too high. high! Press the OK key to retry. Pneumatic The cycler’s Warning: pneumatic pressure Pressure reading is reading too low. low! Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 238 Self Test The self test has Press the OK key to retry. Warning: failed due to a Pressure leak! possible air pressure leak. P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 239 When wait for it to reach the set temperature has been the set temperature. reached, it will automatically start the Fill. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 240 Alarm: Pump A error has occurred. Record error code. vs B error! Insert PIN. Clamp lines. Insert patient connector into stay.safe Organizer. Aseptically disconnect from cycler. Turn off cycler and call service. P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 241 Volume error! the pumps are moving. Record error code. Watchdog An unrecoverable Alarm: error has occurred. Aseptically disconnect yourself Watchdog timer from cycler if connected. Error! Turn off cycler and call service. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 242: Different People Drain Differently

    Different People Drain Differently P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 243 CAPD or on another type of cycler, the amount of time that you need to drain may not be the same as before. Your Liberty Select Cycler is designed to drain at a rate of 100 mL per minute for most Drain cycles.
  • Page 244: Power Failure

    Power Failure Turn power back ON P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 245 Switch on the back of your machine ON if it is not already on. Your machine will then go through a series of self-tests before you may choose whether or not to continue with your treatment. (continued) Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 246 Power Failure (continued) Press STOP to mute P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 247  Check that the new sterile stay•safe cap is securely attached to your catheter extension set before removing it from the stay•safe organizer Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 248 When prompted, open the door and remove the cassette  Restart your entire treatment with fresh dialysate bags and a new cassette Note: If your power failure lasts longer than eight hours, you may not resume your treatment. P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 249 Cleaning, Specifications, Index, and Glossary Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 250: Cleaning Your Cycler

    Cleaning Your Cycler Heater Tray Front Panel Apparatus Cabinet Outer Door Damp Cloth P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 251 Cleaning your cycler The Liberty Select Cycler is not sterile equipment, however, you should keep your Liberty Select Cycler clean and dry at all times. Clean the exterior surface of your machine using a damp cloth and a diluted (1:100) bleach solution or other suitable hospital disinfectant.
  • Page 252: Machine Specifications

    IEC 60601-1-1. If in doubt, consult the Technical Support department at 1-800-227-2572 or your local representative. P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 253 Minimum 25°C bag temperature in order to start treatment Maximum Total 30,000 mL Volume Drain Time 20-59 minutes Fill Volume 500-4000 mL Pause Volume 0-4000 mL Last Fill Volume 0-4000 mL Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 254 Last bag option Drain exit % Language SI Units Personal data USB enabled Add. drain Add. drain alert Flow alert Flow alert sound Accessories IQdrive (P/N 180202) Liberty Cart (P/N 180080) P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 255 WITH RESPECT TO ELECTRICAL SHOCK, FIRE AND MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH UL 60601-1 and IEC 60601-2-39 CAN/CSA – C22.2 No. 601 CAN/CSA – C22.2 No. 601.2.39 Control No. 36CT Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 256: Manufacturer's Emc Declaration

    CISPR 11 low-voltage power supply network that supplies buildings used for domestic purposes. Harmonic Class A emissions IEC 61000- Complies Voltage fluctuations/ flicker emissions IEC 61000- P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 257 IEC 61000-4-8 purposes. NOTE is the a.c. mains voltage prior to application of the test level. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 258 Liberty Cycler. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m. P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 259 Caution: In order to prevent potentially hazardous conditions and ensure the safe operation of the Liberty Select Cycler, Radio Frequency Identification (RFID) Equipment must be kept at least 3 meters from the system while in operation.
  • Page 260: Product Improvement Policy

    Product Improvement Policy Your Liberty Select Cycler machine was designed and built to comply with these product specifications. It is the intention of Fresenius Medical Care to continuously improve products, a process that may result in modifications to specifications or equipment produced in the future.
  • Page 261 In cases where the Liberty Select Cycler is sold, SALE of the machine or parts described or referenced herein is expressly conditioned upon the terms and conditions set forth below. Any additional or different terms or conditions set forth by the Purchaser to Fresenius Medical Care, Inc., (herein called “the Company”) shall not be effective or...
  • Page 262 SHIPPING: The Company will use its own judgment in determining the most effective method of shipping unless otherwise specified by the purchaser. P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 263 For additional servicing and information contact: Fresenius Medical Care (800) 227-2572 Attention: Service Department 4040 Nelson Ave Concord, CA 94520 The Liberty Select Cycler is manufactured by: Fresenius USA, Inc. 4040 Nelson Ave Concord, CA 94520 Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 264: Glossary

    CCPD—Abbreviation for Dialysate—Dialysis solution ‘Continuous Cycling Peritoneal used to clear waste fluid from Dialysis.’ CCPD is a continuous your body. therapy with exchanges occurring throughout the night and P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 265 Panel, keys work outside the cavity. An exchange or cycle Touch Screen to scroll, confirm consists of three steps: Fill, selections, and stop functions. Dwell, and Drain. Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 266 Status Bar from your cycler. It differs from on the Touch Screen. Touching other Dwells because it is not a one of these buttons will display its screen. P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 267 Some parts of during treatment. the supplies must remain sterile after opening the package. Unrecoverable Error—Alarms in which the Liberty Select Systolic pressure—The higher Cycler Display Status Bar turns blood pressure number, it is the red and cannot be resolved;...
  • Page 268: Index

    Electrical plug ......47 Door ........36 Electromagnetic Guide pins ......35 Emissions ......254 Inserting ....104, 106, 107 Environment ......254 Inspecting ......105 Immunity ......254 Leak ........169 P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 269 OK key ........27 Patient button ..... See Buttons STOP key ....... 27 Patient line ..... See Lines Up/Down Arrow keys....27 Patient line (extended) .... 127 Patient settings ......76 Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 270 Extension cap holder ..... 43 Warnings . 206, 207, 212, See Error New cap holder ...... 43 messages Organizer clip ......43 Warranty ........258 Setting up .... 109, 132, 146 Weight P/N 480145 Rev A Liberty Select Cycler User’s Guide...
  • Page 271 After treatment ..... 159, 181 Y  Entering ........82 Yellow line ... See Lines, Drain White line ..See Lines, Solution Liberty Select Cycler User’s Guide P/N 480145 Rev A...
  • Page 272 www.fmcna.com Fresenius USA, Inc. | 4040 Nelson Avenue | Concord, CA 94520...