Fresenius Medical Care Liberty Select User Manual page 235

Peritoneal dialysis cycler
Table of Contents

Advertisement

Message
Meaning
Heater Warning:
The solution
Solution
temperature test
temperature low!
has failed due to
low temperatures.
IQdrive not
The IQdrive is either
detected
not inserted or not
inserted correctly.
IQdrive Port
Your cycler cannot
Error
access the IQdrive.
Make sure the
Your cycler is
heater bag is on
having trouble
the heater tray
pumping dialysate
and the line is
from the heater bag.
not clamped.
Notice: Draining
Your cycler has
slowly
recognized that you
are not draining as
expected.
If "flow alert sound"
option is set to
"YES", you may
hear several beeps
when this message
is displayed. This is
referred to as a
"flow alert". You do
not need to press
STOP key at this
time.
Liberty Select Cycler User's Guide
Action Required
Cold bags will increase warming
time. Be sure to store your bags
at room temperature.
Check if heater plate (tray) is
warm to touch.
Check for external cooling
sources: Be sure there are no
open windows, drafts, or fans or
air conditioners blowing on your
cycler.
Press the OK key to retry.
Remove the IQdrive and firmly
re-insert it to try again, or press
the OK key to ignore the
message and continue without
using the IQdrive.
Turn off cycler, wait 10 seconds,
and turn it back on again. If the
message is not cleared, call
service. If you are not going to
use the IQdrive, press the OK
key to ignore the message and
continue without using the
IQdrive.
Make sure that the Heater Bag
Line (Red clamp) is properly
connected, the clamps are open
and there is no kink in the line.
Make sure that the Heater Bag
is properly positioned on the
Heater Tray (see page 112) and
the cone is broken.
Press the OK key to retry.
Change positions to move fluid
within your abdomen.
Check that your patient line
(blue clamp) is not kinked or
clamped.
Check for fibrin or clots in the
drain line. If any are visible,
contact your PD nurse for
instructions on what to do.
Are you constipated? If so,
contact your PD nurse for
instructions on what to do.
Check that your catheter
extension set is adequately
secured.
P/N 480145 Rev A
233