1 Press STOP
If the add. drain option is set to YES, your cycler will
display the "Select additional drain option" screen after the
last drain is complete.
If add. drain alert option is set to YES, your cycler will
alert you with an alarm sound. Press the STOP key to
silence the audible alarm.
2 Touch YES or NO
YES option:
To perform an additional drain, touch the YES button.
The additional drain will immediately begin draining your
abdomen.
After your additional drain is complete, your cycler will
begin your Last Fill, if programmed, or it will end the
treatment. Your treatment data will show that you
performed an additional drain (see page 176 for more
information).
NO option:
If you do not want to perform an additional drain, touch the
NO button. Your cycler will begin your Last Fill, if
programmed, or it will end the treatment.
Liberty Select Cycler User's Guide
Note: An additional drain works the same way as a
regular Drain. You may stop it or bypass it as
needed.
Note: An additional drain is a planned drain.
STAT drain is an unplanned drain.