Code
M75
M76
M77
M83
MWD
228
Message
Treatment Warning:
Volume error!
Treatment Alarm:
Pause time error!
Treatment Warning:
Encoder flag error!
Treatment Alarm:
Pump A vs B
Volume error!
Watchdog Alarm:
Watchdog timer
Error!
P/N 480145 Rev A
Meaning
Volume is not
being pumped
although the
pumps are
moving.
An unrecoverable
error has
occurred.
Treatment error.
An unrecoverable
error has
occurred.
An unrecoverable
error has
occurred.
Liberty Select Cycler User's Guide
Action Required
Press the OK key
to retry.
Press STOP to
mute.
Record error code.
Insert PIN.
Clamp lines.
Insert patient
connector into
stay.safe
Organizer.
Aseptically
disconnect from
cycler.
Turn off cycler and
call service.
Press the OK key
to retry.
Press STOP to
mute.
Record error code.
Insert PIN.
Clamp lines.
Insert patient
connector into
stay.safe
Organizer.
Aseptically
disconnect from
cycler.
Turn off cycler and
call service.
Record error code.
Aseptically
disconnect
yourself from
cycler if
connected.
Turn off cycler and
call service.