Guidance and manufacturer's declaration – electromagnetic immunity
The Liberty Cycler is intended for use in the electromagnetic environment specified below.
The customer or the user of the Liberty Cycler should assure that it is used in such an
environment.
Immunity test IEC 60601 test
level
Electrostatic
±6 kV contact
discharge
±8 kV air
(ESD)
IEC 61000-4-2
Electrical fast
±2 kV for power
transient/burst
supply lines
±1 kV for
input/output
IEC 61000-4-4
lines
Surge
±1 kV line(s) to
line(s)
±2 kV line(s) to
IEC 61000-4-5
earth
Voltage Dips,
<5 % U
short
(>95 % dip in U
interruptions,
for 0.5 cycles
and voltage
variation on
power supply
40 % U
T
input lines
(60 % dip in U
IEC 61000-4-
for 5 cycles
11
70 % U
T
(30 % dip in U
for 25 cycles
<5 % U
(>95 % dip in U
for 5 seconds
Power-
3 A/m
Frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE
is the a.c. mains voltage prior to application of the test level.
U
T
Liberty Select Cycler User's Guide
Compliance
level
±6 kV contact
±8 kV air
±2 kV for power
supply lines
±1 kV for
input/output lines
±1 kV line(s) to
line(s)
±2 kV line(s) to
earth
<5 % U
T
T
)
(>95 % dip in U
T
for 0.5 cycles
40 % U
T
)
(60 % dip in U
T
for 5 cycles
70 % U
T
)
(30 % dip in U
T
for 25 cycles
<5 % U
T
T
)
(>95 % dip in U
T
for 5 seconds
3 A/m
Electromagnetic environment
– guidance
Floors should be wood, concrete,
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%.
Mains power quality should be
that of a typical low-voltage
power supply network that
supplies hospitals or buildings
used for commercial or domestic
purposes.
Mains power quality should be
that of a typical low-voltage
power supply network that
supplies hospitals or buildings
used for commercial or domestic
purposes.
Mains power quality should be
that of a typical commercial or
)
T
hospital environment.
If the user of the Liberty Cycler
requires continued operation
during power mains interruptions,
it is recommended that the
)
T
Liberty Cycler be powered from a
medical grade uninterruptible
power supply.
)
T
)
T
Power frequency magnetic fields
should be at levels characteristic
of a typical public low-voltage
power supply network that
supplies hospitals or buildings
used for commercial or domestic
purposes.