6 Insert cassette

To insert the cassette:

Make sure that the Liberty Select cycler is positioned
near the front of the table or cart so that all lines can
hang freely from the cassette.
Check the package label on the Liberty cycler set to
ensure it is not past its expiration date, and then remove
the cycler set from its packaging. Inspect the cycler set
to make sure it is free from kinks or damage. Look for
scratches, loose tubing, or small holes in the cassette
film.
Firmly push the door latch to open the cassette door.
Liberty Select Cycler User's Guide
Warning: A damaged cassette can cause peritoneal
dialysis (PD) solution to leak into the cycler and is a
potential risk for contamination.
Notify your PD nurse if you find a fluid leak.
Caution: Use caution when touching the cassette
film to prevent damage.
Warning: DO NOT USE the set if it is damaged.
Put the damaged set aside and immediately call
Customer Service at 800-323-5188 to arrange a
pick-up to return the damaged set.
Caution: DO NOT insert the cassette until the
cycler prompts you with the on-screen instructions.
If you attempt to insert it at any other time, the
cassette will not snap properly into place and the
door will not close.
Caution: Avoid touching any of the sensors inside
the cassette door, as this could cause damage to the
pump module.
Caution: The cassette door is pressurized.
Do not attempt to open the cassette door
until your machine instructs you to do so.