Fresenius Medical Care Liberty Select User Manual page 224

Peritoneal dialysis cycler
Table of Contents

Advertisement

Code
M30
M30.1
M31
M37
M38
222
Message
Self Test Warning:
Balloon Valve leak!
Pneumatic Alarm:
Balloon Valve leak!
Treatment Warning:
Air detected in
cassette!
Self Test Warning:
Patient pressure not
constant!
Treatment Alarm:
Treatment data
error!
P/N 480145 Rev A
Meaning
The self test has
failed due to a
possible balloon
valve leak during
setup.
The test has failed
due to a possible
balloon valve leak
during treatment.
The test failed due
to possible air
detection in
cassette.
The self test has
failed due to an
error with one or
both of the
pressure sensors
used to measure
fluid pressure in
the cassette.
An unrecoverable
error has
occurred.
Liberty Select Cycler User's Guide
Action Required
Press the OK key
to retry.
Press STOP to
mute.
Record error code.
Insert PIN.
Clamp lines.
Insert patient
connector into
stay.safe
Organizer.
Aseptically
disconnect from
cycler.
Turn off cycler and
call service.
Press the OK key
to retry.
Press the OK key
to retry.
Press STOP to
mute.
Record error code.
Insert PIN.
Clamp lines.
Insert patient
connector into
stay.safe
Organizer.
Aseptically
disconnect from
cycler.
Turn off cycler and
call service.