Recovery Failure - Fresenius Medical Care Liberty Select User Manual

Peritoneal dialysis cycler
Table of Contents

Advertisement

(continued)
Press STOP
Press the STOP key to silence the audible warning. Your
cycler will perform a safety systems check. If the Power
Fail Recovery is successful, your machine will beep again.
Press the STOP key to silence the alarm again. Then you
may select the Resume Treatment button to continue your
Drain, Fill, Dwell, or Pause where you left off.
Follow your PD nurse's power failure instructions to
reconnect to your machine if you have disconnected. If
you are instructed to, you may continue with your
treatment.

Recovery failure

If the Power Fail Recovery test fails, you must select the
Cancel Treatment button and start a new treatment.
Follow the instructions on the screen:
Mask, then wash or sanitize hands as instructed
by your PD nurse.
Insert the stay•safe PIN connector connected to
your catheter extension set into the stay•safe
organizer clip below the control panel
Insert a new stay•safe cap on the left side of the
organizer
Turn the blue button on the PIN connector ¼ turn
clockwise and push it in to make sure that a new
PIN is inserted in the line
Close patient line (blue clamp)
Close catheter extension set clamp
Use aseptic technique to carefully disconnect
your catheter extension set from your patient line
and cap off with a new sterile stay•safe cap
Check that the new sterile stay•safe cap is securely
attached to your catheter extension set before removing
it from the stay•safe organizer
Liberty Select Cycler User's Guide
Note: The letter 'F' will be displayed next to your
treatment details in the "My Records" screen to
show a power failure occurred during that
treatment. For more information, see "Treatment
Details" on page 179.
P/N 480145 Rev A
245