Message
Fill complication
encountered.
Please check
heater, patient,
and drain lines.
Heater Alarm:
Solution
temperature
high!
Heater bag not
detected.
Heater Warning:
Heater
temperature too
high!
232
Meaning
Your cycler has
recognized that you
are not filling as
expected.
The solution
temperature test
has failed three
times due to high
temperatures.
Your cycler is
having trouble
pumping dialysate
from the heater bag.
The solution
temperature test
has failed due to
high temperatures.
P/N 480145 Rev A
Action Required
Make sure that the Heater Bag
Line (Red clamp) is properly
connected, the clamps are open
and there is no kink in the line.
Make sure that the Heater Bag
is properly positioned on the
Heater Tray (see page 112).
Make sure that the patient line
(blue clamp) is properly
connected, the clamps are open
and there is no kink in the line.
Make sure that the Drain Line
(Yellow clamp) is properly
connected, the clamps are open
and there is no kink in the line.
Press the OK key to retry.
Press STOP to mute.
Record error code.
Insert PIN.
Clamp lines.
Insert patient connector into
stay.safe Organizer.
Aseptically disconnect from
cycler.
Turn off cycler and call service.
Make sure that the Heater Bag
Line (Red clamp) is properly
connected, the clamps are open
and there is no kink in the line.
Make sure that the Heater Bag
is properly positioned on the
Heater Tray (see page 112) and
the cone is broken.
Press the OK key to retry.
Remove anything from the
heater tray other than the heater
bag.
Make sure cycler is not placed
near a heat source.
Press the OK key to retry.
Liberty Select Cycler User's Guide