Fresenius Medical Care Liberty Select User Manual page 238

Peritoneal dialysis cycler
Table of Contents

Advertisement

Message
Pneumatic
Warning:
Vacuum reading
high!
Pneumatic
Warning:
Vacuum reading
low!
Self Test Alarm:
Fluid
temperature out
of range!
Self Test Alarm:
Safety Clamp
Error!
Self Test
Warning: Balloon
Valve leak!
Self Test
Warning: Patient
pressure not
constant!
Self Test
Warning:
Pressure leak!
236
Meaning
The cycler's
vacuum pressure is
reading too high.
The cycler is
reading low vacuum
pressure.
The solution
temperature test
has failed.
The safety clamp
test has failed three
times.
The self test has
failed due to a
possible balloon
valve leak.
The self test has
failed due to an
error with one or
both of the pressure
sensors used to
measure fluid
pressure in the
cassette.
The self test has
failed due to a
possible air
pressure leak.
P/N 480145 Rev A
Action Required
Press the OK key to retry.
Press the OK key to retry.
Press STOP to mute.
Record error code.
Insert PIN.
Clamp lines.
Insert patient connector into
stay.safe Organizer.
Aseptically disconnect from
cycler.
Turn off cycler and call service.
Press STOP to mute.
Record error code.
If connected, insert PIN.
Clamp lines.
Insert patient connector into
stay.safe Organizer.
Aseptically disconnect from
cycler.
Turn off cycler and call service.
Press the OK key to retry.
Press the OK key to retry.
Press the OK key to retry.
Liberty Select Cycler User's Guide