Fresenius Medical Care Liberty Select User Manual page 241

Peritoneal dialysis cycler
Table of Contents

Advertisement

Message
Meaning
Treatment
An unrecoverable
Alarm:
error has occurred.
Treatment data
error!
Treatment
An unrecoverable
Alarm:
error has occurred.
Treatment start
error!
Treatment
The test failed due
Warning: Air
to possible air
detected in
detection in
cassette!
cassette.
Treatment
Treatment error.
Warning:
Encoder flag
error!
Treatment
Volume is not being
Warning:
pumped although
Volume error!
the pumps are
moving.
Watchdog
An unrecoverable
Alarm:
error has occurred.
Watchdog timer
Error!
Liberty Select Cycler User's Guide
Action Required
Press STOP to mute.
Record error code.
Insert PIN.
Clamp lines.
Insert patient connector into
stay.safe Organizer.
Aseptically disconnect from
cycler.
Turn off cycler and call service.
Press STOP to mute.
Record error code.
Insert PIN.
Clamp lines.
Insert patient connector into
stay.safe Organizer.
Aseptically disconnect from
cycler.
Turn off cycler and call service.
Press the OK key to retry.
Press the OK key to retry.
Press the OK key to retry.
Record error code.
Aseptically disconnect yourself
from cycler if connected.
Turn off cycler and call service.
P/N 480145 Rev A