Code
M69
M70
M71.1
226
Message
Heater Alarm:
Solution
temperature high!
Heater Warning:
Solution
temperature low!
Pneumatic Alarm:
Pressure leak!
P/N 480145 Rev A
Meaning
The solution
temperature test
has failed three
times due to high
temperatures.
The solution
temperature test
has failed due to
low temperatures.
The cycler has
failed due to a
possible air
pressure leak.
Liberty Select Cycler User's Guide
Action Required
Press STOP to
mute.
Record error code.
Insert PIN.
Clamp lines.
Insert patient
connector into
stay.safe
Organizer.
Aseptically
disconnect from
cycler.
Turn off cycler and
call service.
Cold bags will
increase warming
time. Be sure to
store your bags at
room temperature.
Check if heater
plate (tray) is
warm to touch.
Check for external
cooling sources:
Be sure there are
no open windows,
drafts, or fans or
air conditioners
blowing on your
cycler.
Press the OK key
to retry.
Press STOP to
mute.
Record error code.
Insert PIN.
Clamp lines.
Insert patient
connector into
stay.safe
Organizer.
Aseptically
disconnect from
cycler.
Turn off cycler and
call service.