REPLACEMENTS
REPLACEMENTS
REPLACEMENT COMPRESSOR ASSY M9
1
.
Power down ML 700.
2
.
Unplug ML 700.
3
.
Open left SIDE PANEL.
4
.
Take off PROTECTIVE COVER with FILTER. (No FILTER in ML700 Plus).
5
.
Unplug COMPRESSOR M9.
6
.
Unscrew CONNECTOR with PRESSURE TUBING.
7
.
Take out MOUNTING SCREWS.
8
.
Lift front of COMPRESSOR BASE.
PRESSURE TUBING
KODAK AG, Stuttgart
MOUNTING SCREW
figure 2-1
2-3
3053-3
PHILLIPS
SCREWS (4)
COMPRESSOR
COVER
COMPRESSOR
BASE
12/1998