3053-3
CASSETTE CENTRING STOP SWITCHES S14/15 (CCS)
PURPOSE:
The 2 switches S14/15 should be actuated, when there is a clearance of 0.6 - 0.8 mm
between the CENTRING BARS and a CASSETTE.
1
.
Start the Service Mode FDAB,Day,Day,9,9.
2
.
Select Option 7.1( Feed CASSETTE)
3
.
Feed in a CASSETTE.
4
.
Start SENSOR TEST.
5
.
Move CENTRING BARS in by hand.
!
Caution
CHECK THAT THE SWITCHES ARE NOT ACTUATED TOO EARLY. OTHERWISE THEY
MIGHT BE DAMAGED.
6
.
With a clearance of 0.6 - 0.8 mm between CENTRING BARS and a CASSETTE
adjust S14/15 until they are actuated.
12/1998
S 14 /15
figure 4-17
4-16
ADJUSTMENTS
KODAK AG, Stuttgart