12
.
Ensure that the DOUBLE FILM DETECTOR is enabled in the CHANGE
PARAMETER MENU.
PRESS 2 (CHANGE PARAMETER)
Press 8 twice
Press 3 (DOUBLE SHEET DET.)
Press 1 this enables the double film detector
Press 7 (ENABLE OPERATION)
Press 1
Press 8 twice
13.
Mount the REAR PANEL and fix the ML700 to the PROCESSOR (or SIDE-BY-SIDE
SYSTEM).
14.
Run a few cycles with various cassette sizes to ensure proper operation.

TUNNEL SENSOR REAR B7 (TSR)

PURPOSE :
To set the sensitivity of the SENSOR ,to make it give correct output when it is covered by a
FILM.
1
.
Connect positive lead of DVM to Pin 7 of U1 on PCB A9. Connect negative lead of
DVM to TP2 of PCB A9.
TP8
TP7
A9X2
12/1998
14
TP3
Step 2. Without a FILM the voltage should be
TP6
TP5
U1
A9X2A
8
7
6
1
TP4
A9X4
figure 4-82
4-80
ADJUSTMENTS
< +0.4 V
=
+
TP2
-
TP1
-
+
TP9
Step 4. Adjust R28 to + 1.0 V