Replacing The Lower Part Of The Housing (Control Panel Installation) - ProMinent DULCOMETER Assembly And Operating Instructions Manual

Measured variable: chlorine
Table of Contents

Advertisement

Replacing spare part units
14.
Push and lock the ribbon cable (1)
in its base. The catches (2 and 4)
are used to aligned the two halves
of the housing.
15.
Screw the top part of the controller
housing onto the lower part of the
DULCOMETER
troller housing
16.
Re-check that the seal is seated
properly. IP 67 degree of protection
(wall/pipe-mounting) can only be
provided if the installation is correct
1
3
Fig. 45: Suspend and secure the
DULCOMETER
Compact Controller
®
17.
Suspend the DULCOMETER
Compact Controller at the top (1) in
the wall/tube retaining bracket and
push using light pressure at the
bottom (2) against the wall/pipe
retaining bracket. Then press
upwards (3) until the
DULCOMETER
troller audibly snaps into position
92
Compact Con‐
®
2
A0275
®
Compact Con‐
®
14.3
Replacing the lower part
of the housing (control
panel installation)
Complete commissioning of
the controller
Once the lower part of the housing
has been replaced, it is necessary
to fully commission the measuring
and control point, as the new lower
part of the housing does not have
specific settings, only factory set‐
tings.
NOTICE!
Ribbon cable base
The base of the ribbon cable is
firmly soldered onto the PCB. The
base cannot be removed. Open
the base lock (3) to loosen the
ribbon cable, see Fig. 36
Fig. 46: Loosen the ribbon cable from the
base
1.
Undo four screws and open the
DULCOMETER
troller
2.
Open the right and left lock on the
base and pull the ribbon cable (1)
out of the base.
1.
®
Compact Con‐
A0352