Reference Manual
00809-0100-4530
Table I-3: RB.RECOMMENDED_ACTION
Reference Manual
(continued)
Alarm type
Failed
Failed/Maint/Advise Active
event
Level / Interface
Measurement Failure
Volume / Temperature
Measurement Failure
Electronics Failure -
Transducer Block
Probe Missing
Input / Output Failure
Non-Volatile Memory Failure
Resource Transducer Block
April 2019
Recommended action text
string
•
Analyze echo curve for
reason and check device
configuration.
•
Check device physical
installation (for instance
probe contamination).
•
Load default database to
the device and
reconfigure it.
•
If error persists, it might
indicate a hardware
error. Replace the
transmitter head.
•
If Level Measurement
Failure is active, clear
that alert first.
•
Check volume
configuration.
•
Load default database to
the device and
reconfigure it.
•
Check ambient
temperature at
installation site.
•
If error persists, it might
indicate a hardware
error. Replace the
transmitter head.
Replace transmitter head.
Check that the probe is
properly connected.
Replace the transmitter.
•
Load default database to
the device to clear the
error.
•
Download a device
configuration.
•
If error persists, it may
indicate a faulty memory
chip. Replace the
transmitter head.
455