Advanced configuration
April 2019
Procedure
1. In Rosemount Radar Master, select Setup → Tank.
2. Select the Environment tab.
3. Select the Advanced button for more options.
4. In the Vapor Dielectric Constant box, the desired value. You may also use the Vapor
D.5.2
Adjust the lower product dielectric range
If the dielectric constant of the lower product is significantly smaller than the dielectric
constant of water, you may need to adjust the Lower Product Dielectric Range.
Prerequisites
Special adjustments can also be made by trimming the appropriate amplitude thresholds.
See
Example 2: Interface peak not found
Procedure
1. In Rosemount Radar Master, select Setup → Tank.
2. Select the Environment tab.
3. Select the Advanced button for more options.
4. In the Lower Product Dielectric Range list, select the desired range.
404
Dielectric Calculator or the Vapor Dielectric Chart to find the correct value.
Figure D-5: Dielectric Constants Can Be Adjusted in the Tank Environment
Window
for further information.
Rosemount 5300 Level Transmitter
Reference Manual
00809-0100-4530