Emerson 5301HxxxxxxxxxxxxxxZZ Series Reference Manual page 336

Guided wave radar level transmitter
Table of Contents

Advertisement

Specifications and reference data
April 2019
Table A-26: Rosemount 5301 and 5302 Level and/or Interface in Liquids Ordering Information
(continued)
Consolidate to chamber (see
(41)
XC
Consolidate to Chamber
Engineered solutions (see
Rxxxx
Engineered Solutions beyond standard model codes. (Consult factory for details)
Example model string: 5301-H-A-1-S-1-V-1A-M-002-05-AA-I1-M1C1
E-002-05, means 2 ft. and 5 in. probe length. M-002-05, means 2.05 m.
(1)
Not available with explosion-proof, flameproof, or type n approvals.
(2)
Process seal rating. Final rating depends on flange and O-ring selection. See "Temperature and pressure limits" on
Process temperature and pressure
(3)
Requires option None for sealing (no O-ring).
(4)
For other materials, consult the factory.
(5)
Requires Operating Temperature and Pressure code H, P, or C.
(6)
Only available for Probe Type code 3C.
(7)
Consult factory.
(8)
Only available with Operating Temperature and Pressure code S.
(9)
Requires firmware version 2.L3 or later.
(10)
Only available with Operating Temperature and Pressure code H.
(11)
0.79 lb (0.36 kg) standard weight for flexible single lead probe. L=5.5 in. (140 mm). For PTFE covered probes: 2.2 lb (1
kg) standard weight for flexible single lead probe. L=17.1 in. (434 mm).
(12)
Maximum probe length for Duplex 2205 probes is 105 ft (32 m).
(13)
Extra length for fastening is added in factory.
(14)
Requires model 5301.
(15)
Probe weight included if applicable. Give the total probe length in feet and inches or meters and centimeters, depending
on selected probe length unit. If tank height is unknown, please round up to an even length when ordering. Probes can
be cut to exact length in field. Maximum allowable length is determined by process conditions. See
for more probe length guidance.
(16)
Probes are intrinsically safe.
(17)
Requires F
Fieldbus signal output (U
OUNDATION
(18)
G ½ in. SST cable gland is included in delivery.
(19)
Available for tank connection with flange.
(20)
The standard alarm setting is high.
(21)
Weldings in accordance with EN/ISO standards.
(22)
Certificate includes all pressure retaining wetted parts.
(23)
Design and manufacturing according to ASME B31.1. No code stamp available. Welding in accordance with ASME IX.
(24)
Only available with Material of construction code 1, Operating temperature and pressure code S, H, or P, Probe type 3A,
3B, 3V, 4U, 4A, 4B, 4S, 5A, or 5B, together with ASME flanges size 2-in, 3-in, or 4-in.
(25)
Only available with Signal output code H and Probe type code 3V or 4U.
(26)
Suitable for use as a level sensor part of a limiting device, in accordance with EN 12952-11 and EN 12953-9.
(27)
For Probe Type 3A, 3B, 3C, 4A, 4B, 4S, and 4U, and PTFE-coated 5A.
(28)
Transmitters with aluminum housing are not approved for open deck installations.
(29)
Only for Operating Temperature and Pressure code S.
(30)
Not available with PTFE covered probes.
(31)
Consider any temperature limitations dependent on Material of construction, Hazardous locations certifications,
and/or O-ring selection.
(32)
Option code BR5 together with option code QS or QT is pending, consult factory.
(33)
Option code BR5 together with Remote housing options (B1, B2, B3) or Verification reflector options (HL1, HL2, HL3) is
pending, consult factory.
(34)
Not available with option code U1.
336
Rosemount 9901
Chamber)
Engineered
solutions)
rating,
Flange
rating,
parameter listed in
i
Tri-Clamp
rating, and
Plate
design.
Product
certifications).
Reference Manual
00809-0100-4530
Total probe length
Rosemount 5300 Level Transmitter

Hide quick links:

Advertisement

Table of Contents
loading