Emerson 5301HxxxxxxxxxxxxxxZZ Series Reference Manual page 407

Guided wave radar level transmitter
Table of Contents

Advertisement

Reference Manual
00809-0100-4530
D.6.2
Review installation guidelines
The following installation guidelines should be considered when mounting the
transmitter:
Only probe type 3V or 4U must be used. Check the "VC" and "R3" marking on the seal.
Figure D-7: "VC" and "R3" Marking
The maximum probe length is 13 ft. 1 in. (4 m) for probe type 3V and 9 ft. 10 in. (3 m)
for probe type 4U.
Pipe/Chamber is the only supported mounting type. See below for supported pipe
inner diameters:
Probe type
3V
4U
Probe End Projection is disabled when Vapor Compensation is enabled.
Trim Near Zone shall not be used when Dynamic Vapor Compensation is used.
Dynamic Vapor Compensation requires a minimum distance X from the flange to the
surface level to measure the change in the vapor dielectric constant. If the level rises
within this area, the unit switches over to static compensation, using the last known
vapor dielectric constant.
Reference Manual
A. Probe type 3V (coaxial) for 3-4 in. chambers
B. Probe type 4U (rigid single) for 2 in. chambers
Chamber size
3 to 4 in.
2 in.
Advanced configuration
April 2019
Supported pipe inner
diameter
Minimum 2 in. (50 mm)
1.5 to 2.05 in. (38 to 52 mm)
407