April 2019
Table 7-8: Error Messages Displayed on the Rosemount 5300 Display Panel
Error message
FPROM FAIL
HREG FAIL
OMEM FAIL
MWM FAIL
DPLY FAIL
MODEM FAIL
AOUT FAIL
OHW FAIL
ITEMP FAIL
MEAS FAIL
CONFIG FAIL
SW FAIL
For more information on errors, see Errors.
7.4

LED error messages

For Rosemount 5300 Level Transmitters without display, a flashing Light Emitting Diode
(LED) is used for presentation of error messages.
178
Description
An error in the gauge program memory (FPROM) has
been detected during the startup tests.
Note
This resets the gauge automatically.
An error in the transmitter configuration memory
(EEPROM) has been detected. The error is either a
checksum error that can be solved by loading the default
database, or a hardware error.
Note
The default values are used until the problem is solved.
Other memory failure.
An error in the microwave module.
An error in the LCD.
Modem hardware failure.
An error in the Analog Out Module.
An unspecified hardware error has been detected.
An error in the internal temperature measurement.
A serious measurement error has been detected.
At least one configuration parameter is outside the
allowed range.
Note
The default values are used until the problem is solved.
An error has been detected in the transmitter software.
Rosemount 5300 Level Transmitter
Reference Manual
00809-0100-4530
(continued)