7. Check configuration of Probe Type. The following Probe Types are available:
8. Check the entered Probe Length. See
9. Select the TRANSDUCER1100 block, and then select the Geometry tab.
10. Enter the Tank Height. See
11. Select the Mounting Type.
12. If using pipe, chamber or nozzle, then select the Inner Diameter.
142
•
Rigid twin
•
Flexible twin
•
Coaxial, coaxial HP/C, coaxial HTHP
•
Coaxial large, Coaxial large HP/C
•
Rigid single, rigid single HTHP/HP/C, rigid single PTFE
•
Flexible single, flexible single HTHP/HP/C, flexible single PTFE
The transmitter automatically makes some initial calibrations based on the selected
Probe Type. Normally, the Probe Type is pre-configured at the factory.
Probe length
for further information.
Tank height
for further information.