Service and troubleshooting
April 2019
Table 7-2: Error Messages
Message
RAM error
FPROM error
Database (Hreg) error
Microwave Module error
LCD error
Modem error
Analog out error
Internal temperature error
Other hardware error
Measurement error
Configuration error
Software error
172
Description
An error in the gauge data memory
(RAM) has been detected during the
startup tests.
Note
This automatically resets the gauge.
An error in the gauge program memory
(FPROM) has been detected during the
startup tests.
Note
This automatically resets the gauge.
An error in the transmitter configuration
memory (EEPROM) has been detected.
The error is either a checksum error that
can be solved by loading the default
database or a hardware error.
Note
The default values are used until the
problem is solved.
An error in the microwave module.
An error in the LCD is detected.
An error in the modem used for digital
communication has been detected.
An error in the Analog Out Module.
An error in the internal temperature
measurement.
-40 °C<Internal Temperature<85 °C.
An unspecified hardware error has been
detected.
A serious measurement error has been
detected.
At least one configuration parameter is
outside allowed range.
Note
The default values are used until the
problem is solved.
An error has been detected in the
transmitter software.
Reference Manual
00809-0100-4530
Action
Contact Emerson service department.
Contact Emerson service department.
Load default database and restart the
transmitter. Contact Emerson service
department if the problem persists.
Contact Emerson service department.
Contact Emerson service department.
Contact Emerson service department.
Contact Emerson service department.
Contact Emerson service department.
Contact Emerson service department.
Contact Emerson service department.
Load the default database and restart
the transmitter (see
Reset to factory
settings)
Configure the transmitter or upload a
backup configuration file (see
a backup configuration to
Contact Emerson service department
if the problem persists
Contact Emerson service department.
Rosemount 5300 Level Transmitter
Upload
device)