Reference Manual
00809-0100-4530
Figure 7-21
amplitude peak at the interface between the upper and lower products is not detected in
this case.
Figure 7-21: Echo Curve Plot Indicating that the Interface Threshold for the Interface
Peak Is Too High
A. Amplitude
B. Distance
C. Surface Threshold
D. Product Surface Peak
E. The Interface Threshold is above the Interface Peak.
By adjusting Interface Threshold, the peak at the interface between the upper and lower
products is detected as illustrated in
Reference Manual
illustrates a situation where the Interface Threshold is too high. The signal
Figure
7-22.
Service and troubleshooting