Calibrating The Developer Temperature Display Of The M43 And M43A Only - Kodak X-Omat M43 PROCESSOR Service Manual

Hide thumbs Also See for X-Omat M43 PROCESSOR:
Table of Contents

Advertisement

SERVICE MANUAL
Calibrating the DEVELOPER TEMPERATURE DISPLAY of the M43
and M43A Only
[1] Remove the TOP COVER from the PROCESSOR.
[2] Remove the DRIVE SIDE ACCESS PANEL by loosening the 2 SCREWS.
[3] Pull out the SUPPORT SHELF for the ELECTRICAL BOX from the Drive side of the PROCESSOR.
[4] Pull out the ELECTRICAL BOX.
[5] Connect your ESD protective wrist strap to the JACK on the ELECTRICAL BOX.
[6] Open the TOP COVER of the ELECTRICAL BOX. Fasten the COVER open by placing the O-RING on the
PROCESSOR TANK around the POST on the ELECTRICAL BOX COVER.
Pulling Out the ELECTRICAL BOX
Figure 72
O-RING
POST
COVER
ELECTRICAL
BOX
ELECTRICAL
BOX
SUPPORT SHELF
[7] Enter the diagnostic service routine by moving the 5 DIP SWITCHES at S1 on the 100 CIRCUIT BOARD to
the positions outlined in Table 2.
Table 2 DIP SWITCH Positions for Service Routines
Service Routine
No Service Routines Running
DEVELOPER
TEMPERATURE DISPLAY
(M43, M43A only)
84
ESD
JACK
100 CIRCUIT BOARD
SWITCH 1
SWITCH 2
Off
Off
On
On
SWITCH 3
SWITCH 4
Off
Off
Off
Off
September 1995 – 981090
H130_0143HCA
H130_0143HA
SWITCH 5
Off
On