Removing the STATUS INDICATORS

[1] Check that:
• PROCESSOR is de-energized
• main power is disconnected
• TOP COVER is removed
[2] Remove the FEED TRAY by removing:
• 2 SCREWS
• 2 LOCK WASHERS
• 2 WASHERS
Removing the FEED TRAY
Figure 69
[3] Remove the DRAIN VALVE KNOBS by removing the SETSCREWS securing them.
[4] Remove the DRYER TEMPERATURE CONTROL KNOB.
[5] Remove the FRONT ACCESS PANEL by loosening the 2 SCREWS.
[6] Disconnect CONNECTOR P/J12.
[7] Remove the JAM NUT from the malfunctioning INDICATOR LED.
[8] Using PIN EXTRACTION TOOL TL-2105, remove the appropriate PINS from CONNECTOR J12. See the
table.
Table 1 PIN Connections for the STATUS INDICATORS
STATUS INDICATOR
READY INDICATOR DS1
WAIT INDICATOR DS2
SERVICE INDICATOR DS3
[9] Pull the INDICATOR LED to remove the INDICATOR assembly from the DISPLAY PANEL.
[10] Reverse the steps above to install a new STATUS INDICATOR LED.
82
White Wire Connects To:
PIN 1
PIN 3
PIN 5
SCREW (2)
LOCK WASHER (2)
WASHER (2)
FEED TRAY
Red Wire Connects To:
PIN 2
PIN 4
PIN 6
September 1995 – 981090
H130_0084BCA
H130_0084BA