Plastic ROLLERS of the DEVELOPER and FIXER RACK
[1] Check that:
• PROCESSOR is de-energized
• TOP COVER is removed
[2] Remove the RACK from the PROCESSOR. To prevent contamination of the processing solutions, use the
DRIP TRAY and the SPLASH GUARD (FIXER RACK only).
[3] Rinse the RACK with warm water. Using a clean, damp, lint-free cloth, wipe the ROLLERS to remove any
chemical residue.
[4] If you are removing ROLLERS from the DEVELOPER RACK, remove the COVER from the RACK.
[5] Remove the STUD from each end of the ROLLER you wish to remove. Insert a SCREWDRIVER in the notch
at the base of the STUD and pry the STUD loose. Check the STUDS for damage. If necessary, install new
STUDS.
[6] Remove the ROLLER from the RACK.
[7] Using TWEEZERS or a HEMOSTAT, press the 2 TABS on the ROLLER. Remove the GEAR.
[8] Reverse the steps above to install a new ROLLER.
Important
When installing the GEAR onto the ROLLER check that:
• notches in the GEAR align with the POSTS on the SHAFT
• TABS on the ROLLER snap into position.
[9] Check that the GEARS of the RACK mesh and that all ROLLERS turn freely.
Plastic ROLLERS of the DEVELOPER and FIXER RACKS — DEVELOPER RACK Shown
Figure 8
STUD
981090 – September 1995
plastic
ROLLER
GEAR
notch
RACKS
STUD
H130_0173HCA
H130_0173HA
15