Environmental Requirements; Temperature; Relative Humidity; Altitude - Kodak CARESTREAM DRYVIEW 5850 User Manual

Laser imager
Hide thumbs Also See for CARESTREAM DRYVIEW 5850:
Table of Contents

Advertisement

Environmental Requirements

Temperature

Relative Humidity

Altitude

Surface Condition

Environmental Effects

Laser Specifications

February 27, 2009
Operating: 15° to 33° C (59° to 91° F)
Storage: -40° to 60° C (-40° to 140° F)
Operating: 20% to 80% RH, noncondensing
Storage: 10% to 90% RH, noncondensing
30 m (100 ft.) below sea level to 2438 m (8,000 ft.) above sea level
The surface must be a level (must be level within 1°) table top or counter
capable of supporting the weight of the Imager. For the Imager to operate
with no image degradation, surface vibration levels cannot exceed the
following:
2
Freq. Range
G
/Hz
-7
1-50 Hz
1 x 10
-5
50-200 Hz
6 x 10
-6
200-650 Hz
5 x 10
-6
650-1000 Hz
1 x 10
Heat Dissipation: total heat dissipation for the Imager is 883
BTU/hour
Acoustical Noise:
– Less than or equal to 75 dB-A at 1 m during a sound burst of 1
sec. or less
– Less than or equal to 65 dB-A at 1 m during normal operation
– Less than or equal to 50 dB-A at 1 m during idle mode
Wavelength - 810 nm
Power - 50 mW
Radiation class - 3B
Type - Diode
2G0733
Direction
Vertical or horizontal
Vertical or horizontal
Vertical or horizontal
Vertical or horizontal
Specifications