SERVICE MANUAL
Section 6: Electrical

Removing the 100 CIRCUIT BOARD

Important
Only qualified service personnel should service the PROCESSOR.
Important
All the procedures in this section require that you de-energize the PROCESSOR, turn off the main power at the wall
connection, and remove the TOP COVER before beginning the first step of the service procedure.
Warning
Use ESD precautions.
[1] Check that:
• PROCESSOR is de-energized
• main power is disconnected
• TOP COVER is removed
[2] Remove the DRIVE SIDE ACCESS PANEL by loosening the 2 SCREWS.
[3] Pull out the SUPPORT SHELF for the ELECTRICAL BOX from the Drive side of the PROCESSOR.
[4] Pull out the ELECTRICAL BOX.
[5] Connect your ESD protective wrist strap to the JACK on the ELECTRICAL BOX.
[6] Open the TOP COVER of the ELECTRICAL BOX. Fasten the COVER open by placing the O-RING on the
PROCESSOR TANK around the POST on the ELECTRICAL BOX COVER.
78
September 1995 – 981090