SERVICE MANUAL

Removing the HEAT EXCHANGER

Note
Only the M43 and M43A PROCESSORS have a HEAT EXCHANGER in the FIXER TANK. In the Clinic 1
PROCESSOR there are 2 PLUGS in the FIXER TANK instead of a HEAT EXCHANGER. If the PLUGS leak,
remove them and install new PLUGS by following Steps 1 through 4 and Steps 9 and 10 below for removing the
HEAT EXCHANGER.
[1] Check that:
• PROCESSOR is de-energized
• main power is disconnected
• TOP COVER is removed
[2] Drain the DEVELOPER TANK and the FIXER TANK.
[3] Shut off the WATER SUPPLY.
[4] Remove the DEVELOPER RACK or the FIXER RACK, as necessary.
Important
Only do the next 4 steps if you are removing the DEVELOPER HEAT EXCHANGER. The following steps provide
instructions on how to access the DEVELOPER HEAT EXCHANGER.
[5] Remove the FEED TRAY by removing:
• 2 SCREWS
• 2 LOCK WASHERS
• 2 WASHERS
Removing the FEED TRAY
Figure 60
72
SCREW (2)
LOCK WASHER (2)
WASHER (2)
FEED TRAY
September 1995 – 981090
H130_0084BCA
H130_0084BA