Removing The Main Drive Shaft - Kodak X-Omat M43 PROCESSOR Service Manual

Hide thumbs Also See for X-Omat M43 PROCESSOR:
Table of Contents

Advertisement

SERVICE MANUAL

Removing the MAIN DRIVE SHAFT

[1] Before removing the MAIN DRIVE SHAFT, you need to remove the DRIVE MOTOR. Do Steps 1 through 6
on Page 46.
[2] Lift the MAIN DRIVE SHAFT from the SUPPORTS on the PROCESSOR frame. Do not allow the BEARING
to slide off of the SHAFT.
[3] Lift the CHAIN off the SPROCKET on the MAIN DRIVE SHAFT. See Figure 39 on Page 46.
[4] Remove the BEARING from the Drive end of the MAIN DRIVE SHAFT.
[5] Remove the SPROCKET by loosening the SETSCREW and sliding the SPROCKET off the MAIN DRIVE
SHAFT.
[6] Using a ROLL PIN PUNCH TL-1388, remove the PIN securing the WORM, as needed. The WORMS are
identical.
[7] Remove worn or broken parts from the MAIN DRIVE SHAFT, as needed.
[8] If you need to install an entire new MAIN DRIVE SHAFT, remove and inspect all the parts mounted to the
existing SHAFT. Install the parts onto the new MAIN DRIVE SHAFT.
Removing the WORMS and BEARINGS
Figure 40
FLAT SIDE
of BEARING
Drive
End
48
SETSCREW
SPROCKET
MAIN DRIVE SHAFT
THRUST WASHER (2)
THRUST BEARING
COLLAR
FLAT
SIDE
BEARING
TL-1388
ROLL
PIN
WORM
H130_0061BCA
H130_0061BA
September 1995 – 981090