Removing The Water Input Solenoid Valve (L1) - Kodak X-Omat M43 PROCESSOR Service Manual

Hide thumbs Also See for X-Omat M43 PROCESSOR:
Table of Contents

Advertisement

SERVICE MANUAL

Removing the WATER INPUT SOLENOID VALVE (L1)

[1] Check that:
• PROCESSOR is de-energized
• main power is disconnected
• TOP COVER is removed
• main water supply is turned off
Note
Some water may leak. Be ready to wipe up any spilled water.
[2] To minimize the water that may leak, loosen the connection for the INCOMING WATER HOSE at the wall.
[3] Disconnect the water supply to the WATER INPUT SOLENOID.
Disconnecting the Water Supply to the WATER INPUT SOLENOID
Figure 57
INPUT
WATER
SOLENOID
68
DEVELOPER DRAIN HOSE
FIXER DRAIN HOSE
WASH DRAIN HOSE
INCOMING WATER HOSE
H130_0198BCA
H130_0198BA
September 1995 – 981090