Processors - Kodak X-Omat M43 PROCESSOR Service Manual

Hide thumbs Also See for X-Omat M43 PROCESSOR:
Table of Contents

Advertisement

SERVICE MANUAL
Removing the DIVERTER VALVE (L2) and (L3) in the M43 and M43A

PROCESSORS

Note
See Figure 59 on Page 71 when doing this procedure.
[1] Check that:
• PROCESSOR is de-energized
• main power is disconnected
• TOP COVER is removed
• WASH TANK is drained
• main water supply is turned off
[2] Remove the NON-DRIVE SIDE ACCESS PANEL by loosening the 2 SCREWS.
[3] Install CLAMPS TL-2170 onto the 3 HOSES to stop the flow of water.
Note
Some water may leak. Be ready to wipe up any spilled water.
[4] Identify the output HOSES (Normal Operation and Cooling Mode).
[5] Remove the 3 CLAMPS securing the 3 HOSES to the VALVE and remove the HOSES.
[6] Remove the SCREW securing the DIVERTER VALVE to the PROCESSOR frame.
[7] Disconnect the 4 WIRE PLUGS from the DIVERTER VALVE.
[8] Record the position of the wires.
[9] Remove the DIVERTER VALVE from the PROCESSOR.
[10] Install the DIVERTER VALVE by reversing the steps above.
[11] Check that the following parts were installed correctly:
• 4 WIRE PLUGS
• 2 output HOSES
• 1 input HOSE
[12] Use the Service Routine to check for the correct operation of the DIVERTER VALVE. See the Diagnostics
Manual, Publication Number 981093.
70
September 1995 – 981090

Advertisement

Table of Contents
loading

This manual is also suitable for:

X-omat m43a processorX-omat clinic 1 processor