Removing the RECIRCULATION PUMPS and the REPLENISHMENT
PUMP

Removing the DEVELOPER (B5) and FIXER (B6) RECIRCULATION PUMPS

[1] Check that:
• PROCESSOR is de-energized
• main power is disconnected
• TANKS are drained as necessary
• TOP COVER is removed
[2] Remove:
• DRYER RACK
• EVAPORATION COVER
• WASH RACK
[3] Remove the BACK ACCESS PANEL and both SIDE ACCESS PANELS by loosening the 2 SCREWS
securing each of them to the PROCESSOR.
Note
See Figures 45 and 46 on Page 56 when doing this procedure.
[4] Check that you have completed all the steps on Page 56 before you begin this procedure.
[5] Either drain the processing solution from the PROCESSOR TANK whose PUMP you are removing, or install
CLAMPS TL-2170 onto both HOSES connected to the PUMP you are removing in order to stop the flow of
processing solution.
[6] To gain access to the DEVELOPER RECIRCULATION PUMP, slide out the ELECTRICAL BOX from the
Drive side of the PROCESSOR.
[7] Disconnect the PUMP you are removing.
• CONNECTOR P/J2 for the DEVELOPER RECIRCULATION PUMP
• CONNECTOR P/J3 for the FIXER RECIRCULATION PUMP
Note
The DEVELOPER RECIRCULATION PUMP is the PUMP located closest to the Drive side of the PROCESSOR.
The FIXER RECIRCULATION PUMP is the PUMP located closest to the Non-Drive side of the PROCESSOR.
See the figures on Page 56 for the location of the RECIRCULATION PUMPS.
[8] Use EXTENSION TOOL TL-4571 to loosen the 2 SCREW CLAMPS. See Figures 43, 44, 45, and 46 for the
easiest access to the SCREW CLAMPS.
[9] Disconnect the HOSES.
[10] Loosen the SCREW, LOCK WASHER, and WASHER securing the RECIRCULATION PUMP to the frame
of the PROCESSOR.
[11] Install the RECIRCULATION PUMP by reversing the steps above.
[12] Remove the CLAMPS if they were installed in Step 5.
[13] If you drained the PROCESSOR TANK in Step 5, fill the TANK with water until the water level is 7.6 cm
(3 in.) deep.
(a) Check that the HOSES connected to the RECIRCULATION PUMP fill with water.
(b) Check all HOSE connections for leaks.
[14] Check that the TANKS are full of solution.
[15] Energize the RECIRCULATION PUMP by inserting INTERLOCK TOOL TL-4740 into the AC
INTERLOCK SWITCH S5, located near the DRYER ASSEMBLY on the drive side of the PROCESSOR.
[16] Place a MAGNET on the DC INTERLOCK SWITCH S6 on the DRYER PLENUM BRACKET.