OPERATION
..........................................................................................................................................................................................................

Deleting an event

52
The events are mostly alarms (when the measuring or computational result is being affected), warn-
ings or other events such as the opening of the calibration switch.
..........................................................................................................................................................................................................
1. The first line of a screen (info line) indicates if an
event has occurred. The message(s) it-
self/themselves can be seen in the overview of the
"Operating data" ("OPER.") book. Here all events are
outputted by number in line 2.
2. If the "Event acknowledged" message is displayed in
the first line, you can delete the messages by press-
ing Enter provided that the cursor is located in line 2
and one of the calibration or user codes has been
entered beforehand.