Fault and Event Display (coordinate A09)

In the TERZ 94/94-S there are 16 error messages.
The error numbers are indicated in a roll mode, if more than one error is active, the display changes
each second to the next active error (1... 5... 1... 5...).
Number "0" means faultless operation.
Fault
Fault description
No.
0
No fault
1
wrong EEPROM version, invalid constant TERZ 94 PCB board must be
2
Pulse fault: (sensor loss / 10:10000
(X:Y))
3
EEProm Hardware fault
4
5
Current output min/max fault:
(wrong range values)
6
7
8
Pulse output LF: (more than 500 NF
pulses accumulated in the memory)
9
10
Configuration ist wrong
11
12
13
14
15
16
...........................................................................................................................................................................................................
Manual TERZ 94 · EN09a · 2017-11
-----
-----
-----
-----
-----
-----
-----
-----
-----
-----
Remedy
-----
exchanged
Check the electrical connection of
the sensor. Possibly the sensor
must be exchanged.
TERZ 94 PCB board must be
exchanged
Entered range values (C02 and C03)
are wrong or the value of the
counter factor is wrong.
Check the values and chenge them
if necessary.
Value of the LF output pulse factor
(A07) is too high and must be
adapted.
Selected mode of operation is
wrong. Change the operation mode
in coordinate A8.8 to 2.