OPERATION
..........................................................................................................................................................................................................
46
..........................................................................................................................................................................................................
6. The first line of the input mask shows the old value.
In the third line, you can change the old value using
the " " or " " cursor key for the digit and " " or " "
for the input position. Complete your entry by press-
ing Enter. Line 4 shows the permissible input range,
and line 5 the current time and date.
7. If the new value is accepted, the old value in line 1
will be replaced by the new value. Press Esc to close
the input mask.
8. First jump back by one level with Esc. Then press " "
and Enter to jump to the default value. It is used as a
replacement value if a fault occurs or as a fixed val-
ue if it is set appropriately in the operating mode of
the pressure transmitter.
9. Under "Parameters" the data (Upmin, Upmax,
%Upmin, %Upmax) are to be found. They depend on
the pressure transmitter and have to be changed
when the pressure transmitter is replaced. Other ad-
justed values of the pressure transmitter input were
specified in the factory and are not to be changed
when the pressure transmitter is replaced.
In addition, you will have to adjust the type of the
pressure transmitter in the ID display ("TYPE" book)
in the "Pressure transm." chapter (see white box on
the left). Fields p< and p> will be adjusted automati-
cally when the type of the pressure transmitter is
changed