OPERATION
..........................................................................................................................................................................................................

Preconditions for important operations

Setting totalizers

Book: TOT, menu: Totalizer/Setting totalizers/
Set VmL:
Set Vb:
Set Vm:
Set VO:
40

Resetting totalizers

Book: TOT, menu: Totalizer/Totalizer mode/
Reset:
for {Dist. total.|Main totaliz.|Totalizers}
(Totalizers: Resetting main and disturbance totalizers).

Resetting archives

Book: TARIFF, menu: Archive modes/
Archive reset: C code
for {All arch|Last|Event|Dist|Months|Days|Period}.

Resetting logbooks

Book: TARIFF, menu: Logbook modes/
Logbook reset: C code
for {All logb|Event log|Param log}
(All logb: resetting all logbooks except for the calibration log).
For {Cal log}:

Changing the U1 code (Proceed in the same way to change the U2 code.)

Book: OPER., menu: CodeU1/
Enter old U1 code (e.g. 55555555)
Enter new U1 code (e.g. 12345678)

Changing the C1 code (Proceed in the same way to change the C2 and C3 codes.)

Book: OPER., menu: CodeC1/
Enter old C1 code (e.g. 11111111)
Enter new C1 code (e.g. 99999997)
The permissible numeric range for all codes is from 11111111 to 99999998. Numbers, which are
lower than 11111111 (e.g. 10111111), are not permissible.
..........................................................................................................................................................................................................
U code
C code and calibration switch
C code
C code
C code
Calibration switch
User lock is open.
New U1 code is now valid.
Calibration lock is open.
New C1 code is now valid.