Table 1- 1: Test Equipment; Output - Tektronix TDS 430A Technical Reference

Digitizing oscilloscopes
Hide thumbs Also See for TDS 430A:
Table of Contents

Advertisement

Performance Tests
Equipment Required

Table 1- 1: Test Equipment

Item Number and
Description
1.
Attenuator,10X
(three required)
2.
Attenuator, 5X
(two required)
Termination, 50 Ω
3.
Termination, 75 Ω
4.
5.
Cable, Precision Coaxial
(three required)
6.
Cable, Coaxial
7.
Connector, Dual-Banana
(two required)
8.
Connector, BNC "T"
9.
Coupler, Dual-Input
10. Generator, DC Calibra-
tion
11. Generator, Sine Wave
12. Meter, Level and Power
Sensor
13. Splitter, Power
14. Generator, Function
15. Adapter (four required)
1- 16
These procedures use external, traceable signal sources to directly check
warranted characteristics. The test equipment required is shown in Table 1- -1.
Minimum Requirements
Ratio: 10X; impedance 50 Ω; connec-
tors: female BNC input, male BNC

output

Ratio: 5X; impedance 50 Ω; connec-
tors: female BNC input, male BNC
output
Impedance 50 Ω; connectors: female
BNC input, male BNC output
Impedance 75 Ω; connectors: female
BNC input, male BNC output
50 Ω, 36 in, male to male BNC
connectors
75 Ω, 36 in, male to male BNC
connectors
Female BNC to dual banana
Male BNC to dual female BNC
Female BNC to dual male BNC
Variable amplitude to ±110 V; accura-
cy to 0.1%
100 kHz to at least 400 MHz. Variable
amplitude from 12 mV to 2 V
p-p
Frequency accuracy >2.0%
Frequency range:10 MHz to 400MHz.
Amplitude range: 6 mVp-p to 2 V
Frequency range: DC to 1 GHz.
Tracking: >2.0%
Frequency range 5 MHz to 10 MHz.
Square wave transition time ≤25 ns.
Amplitude range: 0 to 10 V
p-p
into 50Ω
Male N to female BNC
TDS 420A, TDS 430A & TDS 460A Performance Verification and Specifications
Example
Tektronix part number
011-0059-02
Tektronix part number
011-0060-02
Tektronix part number
011-0049-01
Tektronix part number
011-0102-01
Tektronix part number
012-0482-00
Tektronix part number
012-1338-00
Tektronix part number
103-0090-00
Tektronix part number
103-0030-00
Tektronix part number
067-0525-02
Wavetek 9100 (or, optionally,
Data Precision 8200, with
1 kV option installed)
Rohde & Schwarz SMY
.
Rohde & Schwarz URV 35,
with NRV-Z8 power sensor
p-p
Rohde & Schwarz RVZ
Tektronix CFG280
Tektronix 103- - 0045- - 00
Purpose
Signal Attenuation
Signal Attenuation
Checking delay match be-
tween channels
Used to test Video Option 05
equipped oscilloscopes only
Signal Interconnection
Used to test Video Option 05
equipped oscilloscopes only
Various Accuracy Tests
Checking Trigger Sensitivity
Checking Delay Match Be-
tween Channels
Checking DC Offset and
Measurement Accuracy
1
Checking Analog Bandwidth,
Trigger Sensitivity, Sample-
rate, External Clock, and
Delay-Time Accuracy
Checking Analog Bandwidth
and Trigger Sensitivity
Checking Analog Bandwidth
Checking External Clock
Checking Analog Bandwidth

Advertisement

Table of Contents
loading

This manual is also suitable for:

Tds 420aTds 460a

Table of Contents