Restore Comport To Cfgdata - Emerson Rosemount Xstream Series Instruction Manual

Gas analyzer
Hide thumbs Also See for Rosemount Xstream Series:
Table of Contents

Advertisement

X-STREAM

7-6-4-2 Restore COMPort to CfgData

Before starting this procedure connect an
external device (e.g. computer) via the serial
interface to the analyzer.
Setup..
Save-Load..
Save-Load
CfgData>COMport!
COMport>CfgData..
Verify!
COMPort>CfgData
Are you sure?
No!
Yes!
Copying data
- PLEASE WAIT -
Procedure
-COMMAND EXECUTED-
7-52
7-6
Save / Restore Configuration Data Sets
X:078000
(i)
Open a standard terminal software on the
computer with it's serial interface setup the
same way as the analyzer's.
Note!
The analyzer's Modbus protocol will be
disabled during COMPort data transfer.
Starting from the measurement screen press
the DOWN key to open the MAIN MENU, enter
the SETUP and next the SAVE-LOAD.. menu.
If system is setup accordingly access level 3
code must be entered to access this menu.
Now select the "COMPort>CfgData.." line and
press the ENTER key.
A screen appears to confirm the operation:
Select Yes! and after pressing the ENTER key
a new screen shows up showing the current
status.
The instrument now overwrites the currently used
(and possibly changed by operator) analyzer
setup by the data stored in an external device.
The restored data is then called CfgData and
used for operating the analyzer.
When the procedure has finished, the
COMMAND EXECUTED screen shows up.
Emerson Process Management GmbH & Co. OHG
Instruction Manual
HASAxE-IM-HS
05/2006