Fig. 4-5: Socket X2 - Pin Assignment - Emerson Rosemount Xstream Series Instruction Manual

Gas analyzer
Hide thumbs Also See for Rosemount Xstream Series:
Table of Contents

Advertisement

Instruction Manual
HASAxE-IM-HS
07/2006
4-4-1 Installation - X-STREAM GP, X-STREAM GPS
Modbus interface
Specification and driving the interface:
Chapter 7
Note1!
Take care of the special installation
instructions in section 4-5!
Legend:
Pin # / Signal
1 / Common
2 / not used
3 / not used
4 / RXD1
5 / TXD1
4-wire configuration
Legend:
Pin # / Signal
Emerson Process Management GmbH & Co. OHG
1 / Common
2 / RXD
3 / TXD
4 / not used
5 / Common
RS 232 interface
6 / not used
7 / not used
8 / RXD0
9 / TXD0
RS 485 interface

Fig. 4-5: Socket X2 - pin assignment

Note2!
X-STREAM analyzers are to be considered
a DTE (Data Terminal Equipment).
6 / not used
7 / not used
8 / not used
9 / not used
1 / Common
2 / not used
3 / not used
4 / not used
5 / D1
2-wire configuration
X-STREAM
6 / not used
7 / not used
8 / not used
9 / D0
4-9